LC-Series Controller: 1-4 Channel
ENMET
Manual Revision Date – June 13, 2017
P a g e
|
7
Manual Part No. – 80003-050
4.1.1 Power Supply Module
The Power Supply is situated on the LHS of the main PCB within the enclosure and provides the power for the whole system.
The power supply is a standard item and does not require any modifications regardless of the type and quantity of detector
heads being used.
TRACO
®
POWER
FG
AC
(N
)
AC
(L
)
Input
:
100
–
240
VAC
47
-63
Hz
0
.375
A max
Output
:
24
VDC
/
625
mA
TML
15124
Fault relay
Fuse 1
Fuse 2
1
2
3
4
5
6
7
8
9
10
11
12
Common L
3
Alarm
Relay
Power Switch
There are two power source options available to the system
100-240V AC Mains Supply, connectors 1, 2 and 3
18 to 28V DC Supply, connectors 4 and 5
There are two fuses situated on the main PCB. These are:
FS1 (T1.0A) AC supply Fuse
FS2 (T500mA) DC Supply Fuse