EX-5165
ENMET
Manual Revision Date – June 14, 2017
P a g e
|
7
Manual Part No. – 80003-094
4.0 Operation of the EX-5165
It is best to have the
EX-5165
transmitters powered up, operational and stable for several minutes before applying calibration or test
gas to them.
When the
EX-5165
transmitter is first powered up, it goes through a series of momentary screens, which identify the instrument model
number, serial number and software revision. After all of the momentary screens have been displayed, the instrument arrives at the
Main Gas Display showing the gas concentration and unit of measurement.
Depending on transmitter configuration and calibration condition, the furthest right character in the display may flash a letter
indicating the instrument status. Consult
ENMET
Distributor or
ENMET
for additional information.
4.1 Start up
4.1.1 Typical Start Up
When power is supplied to the
EX-5165,
the S/T will display the following sequence of information:
N
OTE
:
Software revision may cause variations of display output.
Example of Display
Function
The instrument: Model
EX-5165
The instrument: Serial Number
The instrument: Software Revision
IF
the right most character is a flashing
W
The instrument is in Warm-up mode
This should last about 1 minute
The Signal Output is held at 4mA during warm-up
The instrument: Normal Display Mode
Measurement of target Gas
EX-5165
75-1256
S/W X.X
0%
0%W