24
EQ-9-HP Operation Manual
DOC-8240 Rev. 3
System status word
Bit #
Definition
0 (LSB)
Fault status: 1 = one or more faults present, 0 = no faults
1
Laser status: 1 = on, 0 = off
2
Lamp status: 1 = on, 0 = off
3
Laser temperature: 1 = overtemperature, 0 = OK
4
Reserved
–
always 0
5
Reserved
–
always 0
6
Lamphead temperature: 1 = overtemperature, 0 = OK
7
Reserved
–
always 0
8
Lamphead interlock: 1 = open, 0 = OK
9
External interlock: 1 = open, 0 = OK
10
Reserved
–
always 0
11
Laser photodiode signal: 1 = low, 0 = OK
12
Lamp photodiode signal: 1 = low, 0 = OK
13
Ignition status: 1 = failed to ignite, 0 = OK
14
Reserved
–
always 0
15
Reserved
–
always 0