![Energetiq LDLS EQ-400 Скачать руководство пользователя страница 23](http://html1.mh-extra.com/html/energetiq/ldls-eq-400/ldls-eq-400_operation-manual_2409950023.webp)
EQ-400 Operation Manual Rev.
Error! Reference source not found.
17
C h a p t e r 3
INSTALLATION
Unpacking
Upon arrival, start by inspecting all parts of the system for completeness and any damage
incurred in shipping. The EQ-400 shipping box should contain:
1) EQ-400 Power Supply Controller unit
1) EQ-400 Lamp House unit
1) Interconnecting cable from Lamp House to Power Supply Controller (9-pin D-
sub).
1) AC Power cable
1) Chiller tubing kit (Optional)
If any part is missing or appears damaged, contact Energetiq immediately. Do not attempt to
substitute any parts. There are no user-serviceable parts inside the EQ-400 Lamp House or
Power Supply Controller unit.
Installation Procedure
See Chapter 2 for interface specifications and facilities requirements.
1.
Place the Power/Control Unit in a standard 19-inch rack or on a stable surface. The
PCU must be placed within 3 meters of the Lamp House due to the length of the
interconnecting cable. See drawing OTD-8198 in Appendix A for dimensional details.
Note: The PCU weighs 18.8 kg (41.5 lb). A two-person
lift is recommended. Use care when moving the PCU not
to bend the fiber optic cable sharply. Minimum bend
radius is 50mm.
2.
Connect the Lamp House unit optical output to the user equipment. The beam
should always be either directly coupled to a fiber optic cable, or enclosed in an
appropriate beam pipe, tube, or enclosed space and purged with nitrogen. Operating
the source without any output target or beam transport is not recommended, and may
lead to unsafe operating conditions. Consult Energetiq for applications information
and suggested configurations.
The Lamp House should be mounted in the orientation shown in Figure 4, with the
output window on the side and the lamp mounting surface oriented horizontally. The
lamp has been factory aligned in this position. Mounting the lamp in a different
orientation will cause the plasma position inside the bulb to shift slightly and may
cause performance variations.
Содержание LDLS EQ-400
Страница 3: ...EU Declaration of Conformity...
Страница 4: ......
Страница 6: ......
Страница 26: ...20 EQ 400 Operation Manual Rev Error Reference source not found Figure 5 Remote Interface Schematic...
Страница 32: ...26 EQ 400 Operation Manual Rev Error Reference source not found...
Страница 34: ...28 EQ 400 Operation Manual Rev Error Reference source not found...
Страница 35: ......