Viomax CAS51D
Installation
Hauser
17
5.4
Flow operation
5.4.1
Flow assembly for clear water and small sample volumes
1
2
3
4
A0013352
14
Sensor with flow assembly
1
Transmitter
2
Viomax CAS51D
3
Flow assembly
4
Sensor holder
Secure holder
Ø 6.5 (0.26)
86
(3.39)
30 (1.18)
100 (3.94)
70 (2.76)
**
70 (2.76)
60 (2.36)
A0013290
15
Dimensions in mm (inch)
** Variable length
1. For the necessary installation dimensions, refer to the drawing. Mount the sensor in
a horizontal position (as shown in the drawing).
Drill the holes for the mounting clamps into a wall or panel.
2. Secure the mounting clamps. The required fastening fixtures (screws, wall plugs) are
not included in the scope of delivery of the kit and must be provided by the customer.
Содержание Viomax CAS51D
Страница 39: ......
Страница 40: ...www addresses endress com ...