Installation
Hauser
13
5
Installation
5.1
Installation conditions
5.1.1
Mounting location
‣
Choose a mounting location that can be easily accessed at a later stage.
‣
Ensure that upright posts and fittings are fully secured and vibration-free.
‣
Select an installation location that produces a typical nitrate concentration / a typical
SAC value for the application in question.
‣
Do not install the sensor above aeration discs. Oxygen bubbles may accumulate at the
cuvette gap and distort the measured value.
5.1.2
Orientation
A0013268
5
Sensor orientation, arrow =
direction of flow
‣
Align the sensor so that the cuvette gap is rinsed by
means of the flow.
Flexdip CYA112 wastewater assembly and Flexdip CYH112 bracket
A0013267
6
Horizontal, fixed installation
The installation angle is 90°.
‣
Align the sensor in such a way that the
cuvette gap is rinsed by means of the
flow.
A0013270
7
Suspended vertically from a chain
The installation angle is 0°. Tried and tested
arrangement when operating in ventilated
zones.
‣
Ensure that the sensor is adequately
cleaned. There must be no buildup on the
optical windows.