Measured error
6)
Nitrate
For 0.1 to 50 mg/l NO
3
-N (2 mm cuvette gap):
2 % of full-scale value above 10 mg/l
±0.2 mg/l below 10 mg/l
For 0.01 to 20 mg/l NO
3
-N (8 mm cuvette gap):
2 % of full-scale value above 2 mg/l
±0.04 mg/l below 2 mg/l
SAC
2 % of full-scale value when measuring with potassium
hydrogen phthalate (KHP) as the standard
Repeatability
6)
Nitrate
At least ±0.2 mg/l NO
3
-N
SAC
0.5 % of end of measuring range (for homogeneous media)
Detection limits
Nitrate
• CAS51D-AAA1
0.003 mg/l NO
3
-N
• CAS51D-AAA2
0.013 mg/l NO
3
-N
SAC
In relation to the standard potassium hydrogen phthalate (KHP):
• CAS51D-AAC1
0.045 mg/l COD
• CAS51D-AAC2
0.3 mg/l COD
• CAS51D-AAC3
1.5 mg/l COD
Detection limits
Nitrate
• CAS51D-AAA1
0.01 mg/l NO
3
-N
• CAS51D-AAA2
0.043 mg/l NO
3
-N
SAC
In relation to the standard potassium hydrogen phthalate (KHP):
• CAS51D-AAC1
0.15 mg/l COD
• CAS51D-AAC2
1.0 mg/l COD
• CAS51D-AAC3
5.0 mg/l COD
Long-term drift
Nitrate
Better than 0.1 mg/l NO
3
-N over one week
6)
The measured error includes all of the sensor's and transmitter's uncertainties (measuring chain). It does not include any uncertainties resulting
from the reference material or from any necessary adjustment.