Turbimax W CUS65
Installation
Hauser
9
3.2.4
Immersion operation
!
Note!
• Observe the required mounting clearance for the installation mode selected. Installing the sensor
in pipes or close to a wall can lead to backscattering resulting in signal increase, mainly with
measurements in low turbidity ranges (<100 FNU).
• Make sure the sensor is completey immersed at changing water levels.
a0008990
Fig. 4:
Installation with retractable assembly
Arrow = flow direction of the medium
Installation angle
α
dependent on sensor version:
• CUS65-A: 80°
• CUS65-B: 90°
• CUS65-C, E: 100°
• CUS65-D: 110°
a
1
a0008992
Fig. 5:
Wall distance with immersion pipe
a0008991
Fig. 6:
Immersion pipe angles
Arrow = flow direction of the medium
• CUS65-A: 45 °
• CUS65-D: 90 °
• CUS65-B, -C, -E: 180 ° (straight)
min. 200 (8)
min.
100
(4)
mm (inch)
45°
90°