Installation
Turbimax W CUS65
10
Hauser
3.3
Installation instructions
3.3.1
Measuring system
A complete measuring system comprises:
• Turbidity transmitter CUM740
• Turbidity sensor Turbimax W CUS65
• Immersion pipe CYY105 or
• Retractable assembly Cleanfit CUA451
• Extension cable (optional)
• Junction box (optional)
3.3.2
Notes on installation
Proceed as follows for complete installation of a measuring point:
1.
Install the retractable (if used) into the process.
2.
Connect the water to the rinsing nozzles (if using assembly with cleaning).
3.
Install and connect the turbidity sensor.
4.
Install the immersion assembly (if used) into the process.
"
Caution!
• Please follow national grounding regulations when using metal assemblies and installation
devices.
• The sensor must be installed in an immersion assembly for immersion operation. Do not use the
sensor suspended freely from the cable.
• Screw the sensor into the assembly in such a way that the cable does not become twisted.
• Avoid large traction (e.g. by tugging) on the cable.
!
Note!
• Select a mounting location which can be easily accessed for later calibrations.
• When installing without an assembly, the sensor optics must be immersed
at least 4 cm
(1.5")
into the medium.
• Please pay special attention to the installation instructions in the Operating Instructions of the
assembly used.
a0008988
Abb. 7:
Measuring system with immersion pipe
1
Turbidity sensor Turbimax W CUS65-C
2
Immersion pipe CYY105
3
Turbidity transmitter CUM740
4
Basin or channel
a0008989
Abb. 8:
Measuring system with retractable assembly
1
Turbidity sensor Turbimax W CUS65-A
2
Turbidity transmitter CUM740
3
Retractable assembly Cleanfit CUA451
4
Pipe
1
2
3
4
1
3
2
4