Turbimax CUS50D

Absorption sensor for turbidity and solids

measurements

BA01846C/07/EN/03.20

71473151

2020-02-10

Содержание Turbimax CUS50D

Страница 1: ...Products Solutions Services Operating Instructions Turbimax CUS50D Absorption sensor for turbidity and solids measurements BA01846C 07 EN 03 20 71473151 2020 02 10...

Страница 2: ......

Страница 3: ...sor 16 5 3 Mounting the compressed air cleaning unit 20 5 4 Post installation check 20 6 Electrical connection 21 6 1 Connecting the sensor 21 6 2 Ensuring the degree of protection 23 6 3 Post connect...

Страница 4: ...situation Failure to avoid the dangerous situation can result in a fatal or serious injury LCAUTION Causes consequences If necessary Consequences of non compliance if applicable Corrective action This...

Страница 5: ...Turbimax CUS50D About this document Endress Hauser 5 1 3 Symbols on the device Symbol Meaning Reference to device documentation...

Страница 6: ...rticularly suited for use in the following applications Turbidity measurements based on the principle of light attenuation turbidimetry in accordance with EN ISO 7027 Absorption measurements in liquid...

Страница 7: ...oducts as defective During operation If faults cannot be rectified products must be taken out of service and protected against unintentional operation 2 5 Product safety 2 5 1 State of the art technol...

Страница 8: ...iption 3 1 Product design The sensor features a sensor head with 2 path lengths of 5 mm 0 2 in and 10 mm 0 39 in 1 2 3 4 A0036825 1 CUS50D sensor head 1 Light sources 10 mm 0 39 in 2 Light sources 5 m...

Страница 9: ...h compressed air cleaning 3 1 1 Measuring principle The sensor operates on the principle of light attenuation in accordance with ISO 7027 and meets the requirements of this standard It is suitable for...

Страница 10: ...ed against impact and moisture The original packaging offers the best protection Make sure to comply with the permitted ambient conditions If you have any questions please contact your supplier or you...

Страница 11: ...f the product by affixing to it the mark 4 4 2 Electromagnetic compatibility Interference emission and interference immunity as per EN 61326 1 2013 EN 61326 2 3 2013 NAMUR NE21 2012 4 4 3 ISO 7027 The...

Страница 12: ...ax CUS50D 12 Endress Hauser 5 Installation 5 1 Installation conditions 5 1 1 Dimensions 82 7 3 26 12 6 20 5 0 8 292 2 11 5 40 1 57 209 5 8 25 24 5 1 NPT3 4 GI 36 1 4 0 5 13 4 0 53 A0036366 3 Dimension...

Страница 13: ...bimax CUS50D Installation Endress Hauser 13 82 7 3 26 12 6 20 5 0 8 292 2 11 5 40 1 57 209 5 8 25 24 5 1 NPT3 4 GI 36 1 4 0 5 13 4 0 53 85 9 3 38 64 2 52 A0036582 4 Dimensions with clamp Dimensions mm...

Страница 14: ...tion Turbimax CUS50D 14 Endress Hauser 61 2 4 64 5 2 5 36 9 1 5 4 7 0 19 3 8 1 5 40 1 6 A0036826 5 Dimensions for compressed air cleaning Dimensions mm in Compressed air cleaning 2 bar 29 psi maximum...

Страница 15: ...ameter must be at least 50 mm 2 in Install the sensor in places with consistent flow conditions The best installation location is in the ascending pipe item 1 A0036370 7 Direction of flow Align the se...

Страница 16: ...nstallation marking align the sensor against the flow direction 5 2 Mounting the sensor 5 2 1 Measuring system A complete measuring system comprises Turbimax CUS50D turbidity sensor Liquiline CM44x mu...

Страница 17: ...annel transmitter 2 Turbimax CUS50D turbidity sensor 3 CUA252 flow assembly 4 Direction of flow 5 2 2 Mounting options Installing with CUA120 flow assembly 2 A0036835 10 Installing with CUA120 flow as...

Страница 18: ...h CUA252 flow assembly The installation angle is 90 The arrow indicates the flow direction it runs from the 10 mm 0 39 in path to the 5 mm 0 2 in path A0036836 12 Installing with CUA262 flow assembly...

Страница 19: ...runs from the 10 mm 0 39 in path to the 5 mm 0 2 in path Installing with CUA451 retractable assembly 2 A0036838 14 Installing with CUA451 retractable assembly The installation angle is 90 The arrow i...

Страница 20: ...a way that air bubbles cannot accumulate on it 5 3 Mounting the compressed air cleaning unit Fit the compressed air cleaning unit onto the sensor head to the end stop The nozzle of the compressed air...

Страница 21: ...erating Instructions and must follow the instructions contained therein Prior to commencing connection work ensure that no voltage is present on any cable 6 1 Connecting the sensor Connect the sensor...

Страница 22: ...o sensor input left or via M12 connector right Connecting the cable shield Cable sample does not necessarily correspond to the original cable supplied 1 2 3 18 Terminated cable 1 Outer shield exposed...

Страница 23: ...ble glands Modules are not fully secured The display is not fully secured risk of moisture entering due to inadequate sealing Loose or insufficiently tightened cables cable ends Conductive cable stran...

Страница 24: ...issioning 7 1 Function check Prior to initial commissioning ensure that The sensor is correctly installed The electrical connection is correct Before commissioning check the chemical compatibility of...

Страница 25: ...erform customer calibrations with up to 10 points Application Formazine Factory calibration for the formazine application is carried out with the formazine turbidity standard Sensor measured values in...

Страница 26: ...ct the application that suits your field of application Model name Application Unit Absorption Absorption measurement in any liquid media dosing of flocculant AU OD Formazine Absorption turbidity meas...

Страница 27: ...l mg l ppm There is also the additional option of selecting User unit as the unit Here any unit name or a character string can be assigned to the Basic unit e g OD for absorption The system can be ca...

Страница 28: ...alibration The measured error between the measured value of the device and the laboratory measured value is too large This is corrected by a 1 point calibration 1 2 3 4 5 6 7 8 1 2 3 4 5 6 x y a b A00...

Страница 29: ...point calibration x Measured value y Target sample value a Factory calibration b Application calibration 1 Select a data record 2 Set 2 different calibration points in the medium and enter the corresp...

Страница 30: ...lation is performed outside the calibrated operational range gray line The calibration curve must be monotonically increasing The following sample values for the calibration of the CUS50D sensor can b...

Страница 31: ...ation process The aim is to achieve the highest possible calibration quality in the shortest possible time frame while taking external conditions into account For calibrations in the field in adverse...

Страница 32: ...8 1 2 3 4 5 6 x y f 1 1 a b A0039329 24 Principle of factor calibration x Measured value y Target sample value a Factory calibration b Factor calibration Offset With the Offset function the measured v...

Страница 33: ...l Cleaning duration Severe fouling with rapid buildup of deposits 5 minutes 10 seconds Low risk of fouling 10 minutes 10 seconds 8 1 4 Signal filter The sensor is fitted with an internal signal filter...

Страница 34: ...ression filter in liquids with a high level of turbidity with a high solids content Air bubbles do not cause the measured value to increase in media of this type and thus cannot be eliminated with the...

Страница 35: ...rection on the sensor head B The direction is specified on the calibration kit 4 A B A0036702 Slide the calibration kit A into the final position until the limit stop B Function check The sensor raw v...

Страница 36: ...it The function check is positive if the deviation is within the permitted tolerances see 34 Reference tool Solid state reference CUS50D kit Tolerance 5 10 If the measured values of a calibration data...

Страница 37: ...ting Solution No display no sensor reaction Power supplied to transmitter Sensor connected correctly Buildup on optical windows Connect mains voltage Connect sensor correctly Clean sensor Display valu...

Страница 38: ...The medium in which measurement takes place 10 1 Maintenance tasks 10 1 1 Clean sensor Sensor fouling can affect the measurement results and even cause a malfunction The sensor must be cleaned regula...

Страница 39: ...ion are required or if the wrong product was ordered or delivered As an ISO certified company and also due to legal regulations Endress Hauser is obliged to follow certain procedures when handling any...

Страница 40: ...62 Weld in flow assembly Product Configurator on the product page www endress com cua262 Technical Information TI01152C Flexdip CYA112 Immersion assembly for water and wastewater Modular assembly syst...

Страница 41: ...anywhere on the ground on the coping stone on the wall or directly onto railings Stainless steel version Product Configurator on the product page www endress com cyh112 Technical Information TI00430C...

Страница 42: ...US50D Connection 6 mm 0 24 in Pressure 1 5 to 2 bar 21 8 to 29 psi Materials POM PE PA 6 6 30 glass fiber Order number 71395617 Compressor For compressed air cleaning 230 V AC order number 71072583 11...

Страница 43: ...0 g l Product loss 0 to 100 1000 Measuring range with solids content For solids the achievable ranges depend very much on the media that are actually present and may differ from the recommended operat...

Страница 44: ...ed for calibration 13 2 3 Drift Working on the basis of electronic controls the sensor is largely free of drifts Formazine drift 0 04 per day for 2000 FAU Absorption drift 0 015 per day for 5 OD 13 2...

Страница 45: ...tion Installation 13 5 2 Weight Cable length Plastic sensor Metal sensor Metal sensor with clamp 3 m 9 84 ft 0 46 kg 1 5 lbs 1 15 kg 2 54 lbs 1 21 kg 2 67 lbs 7 m 23 ft 0 68 kg 1 5 lbs 1 37 kg 3 81 lb...

Страница 46: ...put 43 Installation 12 16 Installation conditions 12 M Maintenance 38 Marine 11 Measuring principle 9 Measuring system 16 Mechanical construction 45 Multipoint calibration 27 N Nameplate 10 O Offset 3...

Страница 47: ......