background image

Operation

Turbimax CUE25 / CUE26

12

Hauser

4.4

Instrument calibration

The instrument was calibrated and tested prior to leaving the factory. Therefore, it is possible to use 
it directly out of the box. However, recalibration of the instrument is recommended to help with 
familiarization and the operation of the instrument and the calibration procedures. In addition, 
recalibration is recommended at least once every three months

2)

.

The instrument requires three standards to be fully calibrated.

4.4.1

Calibration standards

We recommend that the following standards be used during calibration to achieve the full-scale 
accuracy stated in this manual:

• 0.02 NTU calibration standard
• 10.0 NTU calibration standard
• 1000 NTU calibration standard

These primary calibration standards are more stable than Formazin and have a minimum shelf life 
of 12 months. Prior to recalibration, review the expiration dates to ensure that the standards have 
not expired.
If you use Formazin to calibrate the instrument, make sure that a fresh stock suspension of Formazin 
is used to achieve the accuracy quoted for the instrument.

!

Note! 
The 10.0 NTU standard is only guaranteed to hold its value for 12 months when stored in the plastic 
bottle. After pouring it into the glass cuvette, it should be used immediately for calibration and then 
be disposed of. If the 10.0 NTU standard is stored in the glass cuvette for extended periods of time, 
it should be used as a reference standard only.

4.4.2

Indexing calibration cuvettes

To achieve greatest accuracy and account for normal scratches and aberrations in cuvette glass 
when calibrating, we recommend indexing the cuvettes.
Standards and standard kits purchased from Hauser are supplied with indexing rings. 

The following steps allow repeatable indexing of calibration standards:

1.

Press and hold the 

F

 key. 

2.

Slowly rotate the standard inside the optical well, one complete revolution (360°) pausing 
between the increments to allow the display to update.

3.

While rotating the standard slowly, observe the measured turbidity and locate the postion of 
the cuvette having the lowest reading.

4.

With the calibration standard positioned at the location having the lowest turbidity reading, 
install the indexing ring over the cap on the standard so that the pointer of the indexing ring 
aligns with the arrow on the instrument.

When using the standards in future, always insert the standard so that the pointer of the indexing 
ring aligns with the arrow. Slowly rotate the standard back and forth about 5° to find the lowest 
point. The standard is then indexed and ready for use.

2) 

The EPA recommends that on-line turbidimeters be calibrated with a primary standard at least once every three months if they are to be used for EPA reporting.

Содержание Turbimax CUE25

Страница 1: ...Operating Instructions Turbimax CUE25 CUE26 Handheld turbidimeter for field measurement AUTHORIZED DISTRIBUTOR InstrumentsAndControl com Houston Texas USA sales InstrumentsAndControl com 281 609 7170...

Страница 2: ...ere Here you can find the steps for installation of the batteries Operation Page 10 Page 11 ff Use this section to familiarize yourself with the operation of the device This section provides informati...

Страница 3: ...Installation instructions 9 3 5 Post installation check 9 4 Operation 10 4 1 Display and operating elements 10 4 2 Key assignment 10 4 3 Routine operation 11 4 4 Instrument calibration 12 5 Maintenan...

Страница 4: ...int check all the connections for correctness Ensure that electrical cables and hose connections are not damaged Do not operate damaged products and secure them against unintentional commissioning Mar...

Страница 5: ...cause damage to the instrument if ignored Note This symbol indicates important items of information Direct Current DC A terminal at which DC is applied or through which DC flows Alternating Current A...

Страница 6: ...0 02 NTU standard 10 0 NTU standard 1000 NTU standard 2 Empty cuvettes and Kimwipes 1 Operating Instructions BA397C 07 en 1 Instruction card If you have any questions please contact your supplier or...

Страница 7: ...pplier about damage to the delivery contents Keep the damaged products until the matter has been settled Check that the scope of delivery is complete and agrees with your order and the shipping docume...

Страница 8: ...llation Turbimax CUE25 CUE26 8 Endress Hauser 3 3 Installation conditions a0004367 Fig 3 Dimensions CUE25 CUE26 a0004368 Fig 4 Dimensions carry case 70 2 75 165 6 50 48 1 88 mm inch 250 10 216 8 50 mm...

Страница 9: ...ce 2 Remove the two screws pos 1 securing the battery cover pos 2 Lift the cover off 3 Place the supplied four AAA alkaline batteries pos 3 into the battery compartment pos 4 ensuring the correct pola...

Страница 10: ...ay of turbidity levels and user guidance in calibration routine 4 2 Key assignment a0003588 Fig 7 Operating elements 1 key used to turn the Turbimax on or off 2 V W keys used to set numerical values a...

Страница 11: ...he cuvette with the black light shield cap Ensure that the outside of the cuvette is dry clean and free from smudges1 7 Place the cuvette into the instrument and press it down until it snaps fully int...

Страница 12: ...d immediately for calibration and then be disposed of If the 10 0 NTU standard is stored in the glass cuvette for extended periods of time it should be used as a reference standard only 4 4 2 Indexing...

Страница 13: ...5 4 Release the F key to initiate the calibration The display will show a 30 second count down 5 When the instrument has completed calibration on the 1000 NTU it will briefly display 1000 to indicate...

Страница 14: ...l store any of the values calibrated prior to exiting You can cycle through the required calibration points 0 02 NTU 10 NTU 1000 NTU by pressing either the V or W keys to individually calibrate with a...

Страница 15: ...or scratches To clean a cuvette proceed as follows 1 Clean the interior and exterior with a detergent solution 2 Rinse the cuvette thoroughly 8 to 10 times with distilled or deionized water to elimina...

Страница 16: ...e it upside down on a soft surface 2 Remove the two screws pos 1 securing the battery cover pos 2 Lift the cover off 3 Remove the old batteries 4 Place four new AAA batteries pos 3 into the battery co...

Страница 17: ...ssories Endress Hauser 17 6 Accessories 6 1 Calibration solutions Calibration kit CUE25 CUE26 full range 0 02 NTU 10 0 NTU 1000 NTU Order no 51518582 6 2 Cuvettes Sample cuvettes CUE25 CUE26 incl caps...

Страница 18: ...ternal error Contact your responsible service Err icon on display reads CAL Bad standards or standards in wrong order Check the standards and recalibrate the instrument See section Instrument calibrat...

Страница 19: ...red please return it cleaned to the sales center responsible Please use the original packaging if possible 7 6 Disposal The device contains electronic components and must therefore be disposed of in a...

Страница 20: ...ime 14 s Reference temperature 25 C 77 F Resolution 0 01 NTU below 100 NTU 0 1 NTU in the range 100 0 to 999 9 NTU 1 0 NTU in the range 1000 to 1100 NTU Maximum measured error 0 to 500 NTU 500 to 1100...

Страница 21: ...1 875 x 2 75 x 6 50 Instrument in case H x W x L 63 x 250 x 216 mm 2 50 x 10 x 8 50 Weight 1 2 kg 2 7 lbs Materials Instrument housing ABS injection molded Sample cuvette Borosilicate glass Carry cas...

Страница 22: ...Incoming acceptance 7 Input 20 Installation 4 7 9 Instrument calibration 12 K Key assignment 10 M Maintenance 15 Batteries 16 Cuvette 15 Measuring system 7 Mechanical construction 21 N Nameplate 6 O...

Страница 23: ...Turbimax CUE25 CUE26 Endress Hauser 23...

Страница 24: ...www endress com worldwide BA397C 07 en 07 06 FM SGML 6 0 DT 71001162...

Отзывы: