Technical data
Turbimax CUE25 / CUE26
20
Hauser
8
Technical data
8.1
Input
8.2
Power supply
8.3
Performance characteristics
8.4
Environment
8.5
Process
Measured variable
Turbidity
Measuring range
0.01 to 1100 NTU/FNU
Batteries
4 AAA alkaline batteries
(over 5000 tests)
Supply voltage
4 x 1.5 V
Response time
< 14 s
Reference temperature
25 °C (77 °F)
Resolution
0.01 NTU below 100 NTU
0.1 NTU in the range 100.0 to 999.9 NTU
1.0 NTU in the range 1000 to 1100 NTU
Maximum measured error
0 to 500 NTU:
500 to 1100 NTU:
±2 % of reading or 0.01 NTU
±3 % of reading
Repeatability
±2 % of reading
Ingress protection
IP 67 / NEMA 4x
Insulation rating
Double insulated, pollution degree 2
Relative humidity
max. 90%, non-condensing
Process temperature
0 to 50 °C / 32 to 122 °F
Medium temperature range
0 to 50 °C / 32 to 122 °F
Sample size
10 to 15 ml