Maintenance
TOCII CA72TOC
88
Hauser
!
Note!
Analog value output during calibration
During calibration, the last TOC value is transmitted to the analog output and held until the
calibration value has been determined. The calibration value is then transmitted to the analog
output (or to analog output channel 1 in the two-channel version) until a new measured value of
the current sample is determined. During calibration, relay 4 (OPERATION CHECK) is open until
a new measured value is present during operation. If the analog output is used for control purposes,
this OPERATION CHECK signal can be used to declare the analog output as invalid.
7.8.3
Empty volume dosing
Use this service menu to determine the dead volume of pump P2 from the separation chamber to
the end of the capillary.
a0012487
Fig. 49:
Position
Step
Display
Description
1
2
The hose of pump P2 is drained.
PUMPS
CALIBRATION
S E R V I C E
CLEANING
FILTERS
ANALYZER ADJUSTMENT
ANALYZER CALIBRATION
EMPTY VOLUME DOSING
CALIBRATION pH SENSOR
XX
DETERMINE EMPTY VOLUME DOSING
PLEASE WAIT
PUMP CONVEYS BACKWARDS