SS2100 TDLAS Gas Analyzer
XA02750C
Hauser
19
Table 5–1 Fuse specifications
Select the replacement solenoid fuse (F2) based on the
number of solenoids installed on the analyzer.
Drawing
Reference
Voltage
Description
Rating
F1
120
VAC
1 Solenoid, Miniature Fuse,
5 x 20 mm, Time Delay
250 VAC
0.125 A
2 Solenoids, Miniature Fuse,
5 x 20 mm, Time Delay
250 VAC
0.25 A
3 Solenoids, Miniature Fuse,
5 x 20 mm, Time Delay
250 VAC
0.4 A
F2
120
VAC
Miniature Fuse, 5 x 20 mm,
Time Delay
250 VAC
0.8 A
F1
240
VAC
1 Solenoid, Miniature Fuse,
5 x 20 mm, Time Delay
250 VAC
0.63 A
2 Solenoids, Miniature Fuse,
5 x 20 mm, Time Delay
250 VAC
1.25 A
3 Solenoids, Miniature Fuse,
5 x 20 mm, Time Delay
250 VAC
2.0 A
F2
240
VAC
Miniature Fuse, 5 x 20 mm,
Time Delay
250 VAC
1.6 A
F1
24VDC
1 Solenoid, Miniature Fuse,
5 x 20 mm, Time Delay
250 VAC
0.25 A
2 Solenoid, Miniature Fuse,
5 x 20 mm, Time Delay
250 VAC
0.4 A
3 Solenoid, Miniature Fuse,
5 x 20 mm, Time Delay
250 VAC
0.8A
F2
Miniature Fuse, 5 x 20 mm,
Time Delay
250 VAC
0.8 A