XA02750C
SS2100 TDLAS Gas Analyzer
6
Hauser
Table 2–1 SS2100 analyzer specifications
Electrical & Communications
Input Voltages
(Electronics)
1
1. Supply voltage not to exceed
10% of nominal.
2. Application dependent.
120 or 240 VAC ±10%, 50 to 60 Hz, 60 W
(2 power connections) —
Standard
18 to 24 VDC, 1.6 A maximum —
Optional
Input Voltages
(SCS Heater,
Optional)
120 VAC or 240 VAC +/-10%, 50 to 60 Hz single
phase, 200 W
Maximum
Current
2 A maximum at 120 VAC
1 A maximum at 240 VAC
1.6 A maximum at 24 VDC
DO Contact Rating
(Inductive Load)
250 VAC, 3 A NO contact, 1.5 A NC contact
24 VDC, 1 A NO and NC contact
Communication
Analog:
(2) 4-20 mA Isolated, 1200 ohms at
24 VDC maximum (concentration only)
—
Optional
Serial:
RS-232C and Ethernet
Protocol:
Modbus Gould RTU or Daniel RTU or
ASCII
Digital Output
(5): Concentration Alarm, General Fault, Valida
-
tion Fail
, Validation 1 Active
, Validation 2
Active
Application Data
Environmental
Temperature
Range/Sample Cell
Temperature
Range
-20 °C to 50 °C (-4 °F to 122 °F)
-10 °C to 60 °C (14 °F to 140 °F)
- Optional
Environmental
Relative
Humidity
5% to 95%, Non-condensing
Heated SCS
Enclosure
Temperature
50 °C (122 °F)
60 °C (140 °F) —
Optional