Prosonic S FDU91/92/93/95/96
E Hauser
9
#
Warning!
All national guidelines applicable must be observed in explosion hazardous areas.
Nozzle installation
Install the sensor at a height so that the blocking distance BD is not undershot, even at maximum fill level. Use
a pipe nozzle if you cannot maintain the blocking distance in any other way. The interior of the nozzle must
be smooth and may not contain any edges or welded joints. In particular, there should be no burr on the inside
of the tank side nozzle end. Note the specified limits for nozzle diameter and length. To minimise disturbing
factors, we recommend an angled socket edge (ideally 45°).
L00-FDU9xxxx-17-00-00-xx-006
Sensor
α
Application
R (max)
r (max)
FDU91 11°
fluids
10 m
0,96 m
bulk materials
5 m
0,48 m
FDU92
11°
fluids
20 m
1,92 m
bulk materials
10 m
0,96 m
FDU93
4°
fluids
25 m
0,87 m
bulk materials
15 m
0,52 m
FDU95
5°
bulk materials
45 m
1,96 m
FDU96
6°
bulk materials
70 m
3,6 m
Sensor
D [mm]
L [mm]
FDU91
80
< 340
100
< 390
FDU92
150
< 400
FDU93
200
< 520
FDU95
250
< 630
FDU96
300
< 800
L
D
FDU 9x
45°