Prosonic S FDU91/92/93/95/96
4
E Hauser
Input
Measuring range
The effective range of the sensors is dependent on the operating conditions. To estimate the range, proceed as
follows (see also the example):
1.
Determine which of the influences shown in the following table are appropriate for your process.
2.
Add the corresponding attenuation values.
3.
From the total attenuation, use the diagram to calculate the range.
Fluid surface
Attenuation
calm
0 dB
waves
5 ... 10 dB
strong turbulence (e.g. stirrers)
10 ... 20 dB
foaming
ask Hauser
Bulk material surface
Attenuation
hard, rough (e.g. rubble)
40 dB
soft (e.g. peat, dust-covered clinker)
40 ... 60 dB
Dust
Attenuation
no dust formation
0 dB
little dust formation
5 dB
heavy dust formation
5 ... 20 dB
Filling curtain in detection range
Attenuation
none
0 dB
small quantities
5 dB
large quantities
5 ... 20 dB
Temperature difference between
sensor and product surface
Attenuation
to 20 °C
0 dB
to 40 °C
5 ... 10 dB
to 80 °C
10 ... 20 dB
L00-FDU9xxxx-05-00-00-xx-001
A:
Attenuation (dB);
R:
Range (m)
Example
• Silo with rubble:
~ 40dB
• small quantities of
filling curtain:
~ 5dB
• little dust:
~ 5dB
total: ~
50dB
=> Range approx. 8 m
for FDU92
0
20
40
60
80
100
110
45
70
25
20
10
0
FDU 95
FDU 96
FDU 93
FDU 92
FDU 91
A [dB]
R [m]
Содержание Prosonic S FDU91
Страница 21: ...Prosonic S FDU91 92 93 95 96 Endress Hauser 21...