Functions of the auxiliary input
With “PULSED MODE”
Assignment
Pulse at auxiliary input
Function
Remarks
RESET TOTALIZER
•
No pulse
•
Pulse between 3...30 V DC;
at least for the duration of the
set start pulse width.
No function
Totalizer (incl. overflows)
is reset
–
BATCHING
•
No pulse
•
Pulse between 3...30 V DC;
at least for the duration of the
set start pulse width.
•
Another pulse during the filling
procedure (at least for the
duration of the set start pulse
width).
No function
Batching cycle is started.
Batching cycle is stopped.
The “BATCHING” option is only
available, if the function “BATCH
VARIABLE” is set to “VOLUME”
(see page 64).
By deactivating the batching
function, the auxiliary input is
automatically set to “POS. ZERO
RETURN”.
With “LEVEL MODE”
Assignment
Voltage at auxiliary input
Function
Remarks
DUAL RANGE MODE
•
No voltage
•
Voltage of 3...30 V DC
Current output operates with
FULL SCALE 1
Current output operates with
FULL SCALE 2
The “DUAL RANGE MODE”
option is only available, if the
current output is switched on
and the dual range mode is
activated.
If the current output is switched
off or the dual range mode
deactivated, then the auxiliary
input is automatically set to
“POS. ZERO RETURN”.
POS. ZERO RETURN
•
No voltage
•
Voltage of 3...30 V DC
Instrument operates normally.
All output signals are set to “0”
(corresponds to no flow).
No simulation of current or
frequency signals is possible
when “Positive zero return” is
activated.
6 Description of Functions
Promag 33
70
Hauser
Содержание PROMAG 33
Страница 43: ...Promag 33 5 Display and Operation Endress Hauser 43...
Страница 46: ...5 Display and Operation Promag 33 46 Endress Hauser...
Страница 84: ...6 Description of Functions Promag 33 84 Endress Hauser...
Страница 98: ...7 Trouble shooting Maintenance and Repairs Promag 33 98 Endress Hauser...
Страница 106: ...8 Dimensions Promag 33 106 Endress Hauser...
Страница 116: ...9 Technical Data Promag 33 116 Endress Hauser...
Страница 125: ......