Function group
BATCHING
BATCH
VARIABLE
In this function, the batching mode can be activated or switched off.
Note!
If the batching mode is activated, the “BATCHING” Function group is first
shown on the display when entering the operating matrix. The “BATCH
VARIABLE” function then moves into first position within this group.
6
OFF
– VOLUME
BATCH
QUANTITY
This function is used to set the required batching quantity.
Note!
Relay 2 can be assigned as a “BATCH CONTACT” (see page 62).
6
5-digit number with floating decimal point (e.g. 240.00 l)
Unit selection
→
see function “VOLUME UNIT”
Factory setting:
0.0000
[unit]
7
Dispay showing which function is assigned to Relay 2.
BATCH
PREWARN
In this function a
pre-batch quantity can be defined which is used for two-stage
batching cycles (see page 63).
Note!
Relay 1 can be assigned as a “PREBATCH CONTACT” (see page 58).
6
5-digit number with floating decimal point (e.g. 200.00 l)
Unit selection
→
see function “VOLUME UNIT”
Factory setting:
0.0000
[unit]
7
Dispay showing which function is assigned to Relay 1.
Note!
Note!
Note!
6 Description of Functions
Promag 33
64
Hauser
Содержание PROMAG 33
Страница 43: ...Promag 33 5 Display and Operation Endress Hauser 43...
Страница 46: ...5 Display and Operation Promag 33 46 Endress Hauser...
Страница 84: ...6 Description of Functions Promag 33 84 Endress Hauser...
Страница 98: ...7 Trouble shooting Maintenance and Repairs Promag 33 98 Endress Hauser...
Страница 106: ...8 Dimensions Promag 33 106 Endress Hauser...
Страница 116: ...9 Technical Data Promag 33 116 Endress Hauser...
Страница 125: ......