4.3
Connecting the cable of the remote version
Warning!
Danger from electric shock! Switch off the power supply before opening
the instrument.
1. Loosen the safety grip and remove the cover of the
transmitter housing.
2. Remove the cover from the
connection housing of the sensor:
– Promag A, H: Loosen all the Phillips screws
– Promag F:
Loosen the safety grip and unscrew the cover.
3. Feed both signal and coil-current cable into the appropriate cable entries
of the connection housings.
Caution!
Danger of destroying the coil current control! Only connect or disconnect
the coil cable once the power supply to the instrument has been switched off.
4. Connect the sensor / transmitter cable according to the wiring diagrams
(see Fig. 25).
5. Retighten the connection housing cover securely.
With Promag F, the Allen screw of the safety grip also has to be tightened.
Warning!
Caution!
42
41
22
23
36
37
4
8
7
5
6
Transmitter
ba00
9y
31
➊ ➎
➋ ➎
➌
➍
Signal cable
Coil current cable
FS
FL
Coil current
cable
Promag F
Promag H
Sensors
Promag A
Fig. 24
Connecting the transmitter /
sensor cable
Promag 33
4 Electrical Connection
Hauser
25
Содержание PROMAG 33
Страница 43: ...Promag 33 5 Display and Operation Endress Hauser 43...
Страница 46: ...5 Display and Operation Promag 33 46 Endress Hauser...
Страница 84: ...6 Description of Functions Promag 33 84 Endress Hauser...
Страница 98: ...7 Trouble shooting Maintenance and Repairs Promag 33 98 Endress Hauser...
Страница 106: ...8 Dimensions Promag 33 106 Endress Hauser...
Страница 116: ...9 Technical Data Promag 33 116 Endress Hauser...
Страница 125: ......