Trouble-shooting
Proline t-mass 65 PROFIBUS DP/PA
108
Hauser
10.5.2
Replacing the device fuse
#
Warning!
Risk of electric shock. Exposed components carry dangerous voltages. Switch off the power
supply before you remove the cover of the electronics compartment.
The main fuse is on the power unit board.
The procedure for replacing the fuse is as follows:
1.
Switch off the power supply.
2.
3.
Remove the protection cap (1) and replace the device fuse (2).
Only use the following fuse type:
– Power supply 20 to 55 V AC / 16 to 62 V DC
2.0 A slow-blow / 250 V; 5.2 x 20 mm
– Power supply 85 to 260 V AC
0.8 A slow-blow / 250 V; 5.2 x 20 mm
– Ex-rated devices
see the Ex documentation.
4.
Installation is the reverse of the removal procedure.
"
Caution!
Use only original Hauser parts.
A0001148
Fig. 51:
Replacing the device fuse on the power unit board
1
Protective cap
2
Device fuse
1
2