Proline Prosonic Flow P 500
Hauser
71
IS
Transmitter
Sensor
Class I Division 2 Groups A - D
Class I, II, III Division 1 Groups A-G
NI
Transmitter
Sensor
Class I Division 2 Groups A - D
Class I Division 2 Groups A - D
Ex i
Transmitter
Sensor
Class I Zone 2, AEx/Ex nA nC IIC T5…T4 Gc
Class I Zone 1, AEx/Ex d ia IIC T6…T1 Gb
Ex nA
Transmitter
Sensor
Class I Zone 2, AEx/Ex nA nC IIC T5…T4 Gc
Class I Zone 2, AEx/Ex nA ic IIC T6…T1 Gc
Ex tb
Transmitter
Sensor
–
Zone 21, AEx/Ex ia tb IIIC T** °C Db
Functional safety
The measuring device can be used for flow monitoring systems (min., max., range) up to SIL 2
(single-channel architecture; order code for "Additional approval", option LA) and SIL 3 (multi-
channel architecture with homogeneous redundancy) and is independently evaluated and certified in
accordance with IEC 61508.
The following types of monitoring in safety equipment are possible:
Volume flow
Functional Safety Manual with information on the SIL device
HART certification
HART interface
The measuring device is certified and registered by the FieldComm Group. The measuring system
meets all the requirements of the following specifications:
• Certified according to HART 7
• The device can also be operated with certified devices of other manufacturers (interoperability)
Radio approval
The measuring device has radio approval.
For detailed information regarding radio approval, see the Special Documentation → 76
Additional certification
Tests and certificates
• EN10204-3.1 material certificate, parts and sensor housing in contact with medium
• Ambient temperature –50 °C (–58 °F) (order code for "Test, certificate", option JN)
• EN10204-2.1 confirmation of compliance with the order and EN10204-2.2 test report
Other standards and
guidelines
• EN 60529
Degrees of protection provided by enclosures (IP code)
• EN 61010-1
Safety requirements for electrical equipment for measurement, control and laboratory use -
general requirements
• IEC/EN 61326
Emission in accordance with Class A requirements. Electromagnetic compatibility (EMC
requirements).