Installation
Proline Prosonic Flow 93T Portable
8
Hauser
2.2
Preparatory steps prior to installation
Depending on the conditions specific to the measuring point (e.g. Clamp On, number of
traverses, fluid, etc.), a number of preparatory steps have to be taken before actually installing
the sensors:
1.
Using the conditions specific to the measuring point, determine which installation
distances are necessary for installation.
2.
Determine the values for the requisite installation distances.
3.
Mechanical preparation of the Clamp On holders for the sensors:
– Mount the sensor holder (DN 15 to 65 / ½ to 2½")
– Premount the strapping bands (DN 50 to 200 / 2 to 8") or
(DN 250 to 4000 / 10 to 160")
2.3
Determining the necessary installation distances
The installation distances that are necessary for installation depend on:
• The type of sensor: Prosonic Flow P DN 50 to 4000 (2 to 160") or DN 15 to 65 (½ to 2½")
• Number of traverses
2.4
Connecting and switching on the transmitter
A0011547
Fig. 3: Connecting and switching on the transmitter
1
On/off switch (press switch
≥
3 seconds)
2
Charger connection (different adapters are available for the connection)
DN 50 to 4000 (2 to 160")
DN 15 to 65 (½ to 2½")
1 traverse
2 traverses
1 traverse
2 traverses
SENSOR DISTANCE
SENSOR DISTANCE
SENSOR DISTANCE*
WIRE LENGTH
POSITION SENSOR
POSITION SENSOR*
* In the SENSOR DISTANCE function, the distance is indicated in millimeters. The POSITION SENSOR function displays
the values for using the mounting rail (e.g. A3).