Installation
Proline Prosonic Flow 93T Portable
24
Hauser
8.
Screw the connecting cable into the individual sensor.
A0001112
Fig. 21: Mounting the sensor and connecting the connecting cable
This completes the mounting process. The sensors can now be connected to the transmitter via
the connecting cables
→
ä
25.
2.9
Post-installation check
• Are cables or the device damaged (visual inspection)?
• Does the device correspond to specifications at the measuring point, including process
temperature, ambient temperature, minimum fluid conductivity, measuring range, etc.?
• Are the measuring point number and labeling correct (visual inspection)?
• Have the inlet and outlet runs been observed?
• Is the measuring device protected against moisture and direct sunlight?