Proline Promass
Installation
Hauser
15
DN
Ø orifice plate, pipe restriction
[mm]
[in]
[mm]
[in]
150
6
90
3,54
250
10
150
5,91
300
12
210
8,27
350
14
210
8,27
400
16
210
8,27
FB = Full bore
Orientation
The direction of the arrow on the sensor nameplate helps you to install the sensor according
to the flow direction.
Orientation
Recommendation
A
Vertical orientation
A0015591
B
Horizontal orientation, transmitter at
top
A0015589
1)
Exceptions:
→ 3, 16
C
Horizontal orientation, transmitter at
bottom
A0015590
2)
Exceptions:
→ 3, 16
D
Horizontal orientation, transmitter at
side
A0015592
3)
4)
5)
1)
Applications with low process temperatures may decrease the ambient temperature. To maintain the minimum
ambient temperature for the transmitter, this orientation is recommended.
2)
Applications with high process temperatures may increase the ambient temperature. To maintain the maximum
ambient temperature for the transmitter, this orientation is recommended.
3)
Promass A, E, F, G, O
4)
Promass X
5)
Promass H, I, P, Q, S
Содержание Proline Promass series