Proline Promass
Installation
Hauser
29
10 mm
13 mm
1
2
3
4
5
6
7 8
A0019746
10
Post retainer mounting kit
1
8 x hexagonal nut M8 × 0.8
2
4 x threaded bolt M8 × 150
3
1 x post retaining plate
4
1 x post securing plate
5
4 x spring washer M8
6
4 x hexagon bolt M6 × 20
7
4 x spring washer M6
8
4 x hexagonal nut M6 × 0.8
Zero point adjustment
All measuring devices are calibrated in accordance with state-of-the-art technology.
Calibration takes place under reference conditions . Therefore, a zero point adjustment in the
field is generally not required.
Experience shows that zero point adjustment is advisable only in special cases:
• To achieve maximum measuring accuracy even with low flow rates
• Under extreme process or operating conditions (e.g. very high process temperatures or very
high-viscosity fluids).
5.2
Mounting the measuring device
5.2.1
Required tools
For transmitter
• For turning the transmitter housing: Open-ended wrench8 mm
• For opening the securing clamps: Allen key3 mm
• For turning the transmitter housing: Open-ended wrench8 mm
• For opening the securing clamps: Allen key3 mm