For more information on planning and installing EtherNet/IP networks, please refer to the
"Media Planning and Installation Manual. EtherNet/IP" of the ODVA Organization.
Connecting cable between Safety Barrier Promass 100 and measuring device
Cable type
Shielded twisted-pair cable with 2x2 wires. When grounding the cable shield,
observe the grounding concept of the plant.
Maximum cable resistance
2.5 Ω, one side
‣
Comply with the maximum cable resistance specifications to ensure the operational reliability
of the measuring device.
The maximum cable length for individual wire cross-sections is specified in the table below. Observe
the maximum capacitance and inductance per unit length of the cable and connection values for
Wire cross-section
Maximum cable length
[mm
2
]
[AWG]
[m]
[ft]
0.5
20
70
230
0.75
18
100
328
1.0
17
100
328
1.5
16
200
656
2.5
14
300
984
Performance characteristics
Reference operating
conditions
• Error limits based on ISO 11631
• Water with +15 to +45 °C (+59 to +113 °F) at2 to 6 bar (29 to 87 psi)
• Specifications as per calibration protocol
• Accuracy based on accredited calibration rigs that are traced to ISO 17025.
To obtain measured errors, use the
Applicator
Maximum measured error
o.r. = of reading; 1 g/cm³ = 1 kg/l; T = medium temperature
Base accuracy
Mass flow and volume flow (liquids)
±0.10 %
Mass flow (gases)
±0.50 % o.r.
Density (liquids)
• Reference conditions:±0.0005 g/cm³
• Standard density calibration:±0.02 g/cm³
(valid over the entire temperature range and density range )
• Wide-range density specification (order code for "Application package", option EF "Special density
and concentration" or EH " Special density and viscosity"): ±0.004 g/cm³ (valid range for special
density calibration: 0 to 2 g/cm³, +10 to +80 °C (+50 to +176 °F))
Temperature
±0.5 °C ± 0.005 · T °C (±0.9 °F ± 0.003 · (T – 32) °F)