Diagnostics and troubleshooting
Proline Promass F 100 Modbus RS485
82
Diagnostic
number
Short text
Remedy instructions
Status
signal
[from the
factory]
Diagnostic
behavior
[from the
factory]
485
Simulation measured
variable
Deactivate simulation
C
Warning
590
Special event 3
Contact service
F
Alarm
591
Special event 7
Contact service
F
Alarm
592
Special event 11
Contact service
F
Alarm
Diagnostic of process
830
Sensor temperature too
high
Reduce ambient temp. around
the sensor housing
S
Warning
831
Sensor temperature too
low
Increase ambient temp. around
the sensor housing
S
Warning
832
Electronic temperature too
high
Reduce ambient temperature
S
Warning
833
Electronic temperature too
low
Increase ambient temperature
S
Warning
834
Process temperature too
high
Reduce process temperature
S
Warning
835
Process temperature too
low
Increase process temperature
S
Warning
843
Process limit
Check process conditions
S
Warning
862
Partly filled pipe
1. Check for gas in process
2. Adjust detection limits
S
Warning
910
Tubes not oscillating
1. Check electronic
2. Inspect sensor
F
Alarm
912
Medium inhomogeneous
1. Check process cond.
2. Increase system pressure
S
Warning
912
Inhomogeneous
S
Warning
913
Medium unsuitable
1. Check process conditions
2. Check electronic modules or
sensor
S
Alarm
944
Monitoring failed
Check process conditions for
Heartbeat Monitoring
S
Warning
948
Tube damping too high
Check process conditions
S
Warning
990
Special event 4
Contact service
F
Alarm
991
Special event 8
Contact service
F
Alarm
992
Special event 12
Contact service
F
Alarm
1)
Diagnostic behavior can be changed.