MANUAL

BA01659O/06/EN/02.17 

Valid as of version 01.03 zz (Device firmware)

71382654

Operating Instructions Manual

Proline Promass F 100 Coriolis Flowmeter

 

Modbus RS485

Содержание Proline Promass F 100

Страница 1: ...MANUAL BA01659O 06 EN 02 17 Valid as of version 01 03 zz Device firmware 71382654 Operating Instructions Manual Proline Promass F 100 Coriolis Flowmeter Modbus RS485 ...

Страница 2: ...this documentation have been carefully checked and compiled by the author However we cannot completely exclude the possibility of errors TechnipFMC is always grateful to be informed of any errors Contact us on the website Smith Meter is a registered trademark of TechnipFMC Technical Support Contact Information Field Service Response Center 24 7 Technical Support Schedule a Technician 1 844 798 381...

Страница 3: ...1 Installation conditions 20 6 1 1 Mounting position 20 6 1 2 Requirements from environment and process 23 6 1 3 Special mounting instructions 25 6 2 Mounting the measuring device 26 6 2 1 Required tools 26 6 2 2 Preparing the measuring device 26 6 2 3 Mounting the measuring device 27 6 3 Post installation check 27 7 Electrical connection 28 7 1 Connection conditions 28 7 1 1 Required tools 28 7 1...

Страница 4: ... troubleshooting 76 12 2 Diagnostic information via light emitting diodes 77 12 2 1 Transmitter 77 12 2 2 Safety Barrier Promass 100 77 12 3 Diagnostic information in DeviceCare or FieldCare 77 12 3 1 Diagnostic options 77 12 3 2 Calling up remedy information 79 12 4 Diagnostic information via communication interface 79 12 4 1 Reading out diagnostic information 79 12 4 2 Configuring error response...

Страница 5: ... 16 6 Performance characteristics 100 16 7 Installation 105 16 8 Environment 105 16 9 Process 106 16 10 Mechanical construction 109 16 11 Operability 112 16 12 Certificates and approvals 113 16 13 Application packages 114 16 14 Accessories 115 16 15 Supplementary documentation 115 Index 118 ...

Страница 6: ...avoid this situation can result in serious or fatal injury CAUTION CAUTION This symbol alerts you to a dangerous situation Failure to avoid this situation can result in minor or medium injury NOTICE NOTE This symbol contains information on procedures and other facts which do not result in personal injury 1 2 2 Electrical symbols Symbol Meaning Direct current Alternating current Direct current and ...

Страница 7: ...hat are preferred Forbidden Procedures processes or actions that are forbidden Tip Indicates additional information Reference to documentation A Reference to page Reference to graphic Notice or individual step to be observed 1 2 3 Series of steps Result of a step Help in the event of a problem Visual inspection 1 2 5 Symbols in graphics Symbol Meaning 1 2 3 Item numbers 1 2 3 Series of steps A B C...

Страница 8: ...ation Storage and transport Installation Transmitter Brief Operating Instructions Guides you quickly to the 1st measured value Part 2 The Transmitter Brief Operating Instructions are aimed at specialists with responsibility for commissioning configuring and parameterizing the measuring device until the first measured value Product description Installation Electrical connection Operation options Sy...

Страница 9: ...Proline Promass F 100 Modbus RS485 About this document 9 Microsoft Registered trademark of the Microsoft Corporation Redmond Washington USA TRI CLAMP Registered trademark of Ladish Co Inc Kenosha USA ...

Страница 10: ...ce can also measure potentially explosive flammable poisonous and oxidizing media Measuring devices for use in hazardous areas in hygienic applications or where there is an increased risk due to process pressure are labeled accordingly on the nameplate To ensure that the measuring device remains in proper condition for the operation time Keep within the specified pressure and temperature range Onl...

Страница 11: ... In the event of a measuring tube breakage for a device version without rupture disk it is possible for the pressure loading capacity of the sensor housing to be exceeded This can lead to rupture or failure of the sensor housing 2 3 Workplace safety For work on and with the device Wear the required personal protective equipment according to federal national regulations For welding work on the pipi...

Страница 12: ...ements It also complies with the EU directives listed in the device specific EU Declaration of Conformity Endress Hauser confirms this by affixing the CE mark to the device 2 6 IT security We only provide a warranty if the device is installed and used as described in the Operating Instructions The device is equipped with security mechanisms to protect it against any inadvertent changes to the devi...

Страница 13: ...ice is available as a compact version The transmitter and sensor form a mechanical unit 3 1 Product design 3 1 1 Device version with Modbus RS485 communication type 3 2 1 4 A0017609 1 Important components of a measuring device 1 Sensor 2 Transmitter housing 3 Main electronics module 4 Transmitter housing cover In the case of the device version with Modbus RS485 intrinsically safe the Safety Barrie...

Страница 14: ...e Ser no Ext ord cd i i Date Do the nameplate data match the ordering information on the delivery note A0028673 Is the CD ROM with the Technical Documentation depends on device version and documents present If one of the conditions is not satisfied contact your Endress Hauser Sales Center Depending on the device version the CD ROM might not be part of the delivery The Technical Documentation is av...

Страница 15: ...cumentation refer to the following The chapters Additional standard documentation on the device 8 and Supplementary device dependent documentation 8 The W M Device Viewer Enter the serial number from the nameplate www endress com deviceviewer The Endress Hauser Operations App Enter the serial number from the nameplate or scan the 2 D matrix code QR code on the nameplate 4 2 1 Transmitter nameplate...

Страница 16: ...plosion protection Pressure Equipment Directive and degree of protection 8 Flow direction 9 Manufacturing date year month 10 2 D matrix code 11 Document number of safety related supplementary documentation 12 CE mark C Tick 13 Surface roughness 14 Permitted ambient temperature Ta Order code The measuring device is reordered using the order code Extended order code The device type product root and ...

Страница 17: ...ction data e g available inputs and outputs supply voltage 4 Approval information for explosion protection 5 Safety warning 6 Communication specific information 7 Intrinsically safe area 8 Manufacturing location 9 Document number of safety related supplementary documentation 10 Permitted ambient temperature Ta 11 CE mark C Tick 4 2 4 Symbols on measuring device Symbol Meaning WARNING This symbol a...

Страница 18: ...n a dry and dust free place Do not store outdoors Storage temperature 105 5 2 Transporting the product Transport the measuring device to the measuring point in the original packaging A0029252 Do not remove protective covers or caps installed on process connections They prevent mechanical damage to the sealing surfaces and contamination in the measuring tube 5 2 1 Measuring devices without lifting ...

Страница 19: ... a forklift 5 3 Packaging disposal All packaging materials are environmentally friendly and 100 recyclable Measuring device secondary packaging polymer stretch film that conforms to EC Directive 2002 95 EC RoHS Packaging Wood crate treated in accordance with ISPM 15 standard which is confirmed by the affixed IPPC logo or Carton in accordance with European Packaging Directive 94 62EC recyclability ...

Страница 20: ... arising from accumulation of gas bubbles in the measuring tube avoid the following mounting locations in the pipe Highest point of a pipeline Directly upstream of a free pipe outlet in a down pipe Installation in down pipes However the following installation suggestion allows for installation in an open vertical pipeline Pipe restrictions or the use of an orifice with a smaller cross section than...

Страница 21: ...striction mm in mm in 8 6 0 24 15 10 0 40 25 1 14 0 55 40 1 22 0 87 50 2 28 1 10 80 3 50 1 97 100 4 65 2 60 150 6 90 3 54 250 10 150 5 91 Orientation The direction of the arrow on the sensor nameplate helps you to install the sensor according to the flow direction direction of medium flow through the piping Orientation Recommendation A Vertical orientation A0015591 B Horizontal orientation transmi...

Страница 22: ...smitter this orientation is recommended If a sensor is installed horizontally with a curved measuring tube match the position of the sensor to the fluid properties 1 2 A0028774 6 Orientation of sensor with curved measuring tube 1 Avoid this orientation for fluids with entrained solids Risk of solids accumulating 2 Avoid this orientation for outgassing fluids Risk of gas accumulating Inlet and outl...

Страница 23: ...fied gases In suction lines Ensure the system pressure is sufficiently high to prevent cavitation and outgassing For this reason the following mounting locations are recommended At the lowest point in a vertical pipe Downstream from pumps no danger of vacuum A0028777 Thermal insulation In the case of some fluids it is important to keep the heat radiated from the sensor to the transmitter to a low ...

Страница 24: ...eating when heating Ensure that the temperature at the lower end of the transmitter housing does not exceed 80 C 176 F Ensure that convection takes place on a sufficiently large scale at the transmitter neck Ensure that a sufficiently large area of the housing support remains exposed The uncovered part serves as a radiator and protects the electronics from overheating and excessive cooling Heating...

Страница 25: ...not use a heating jacket Make sure that the function and operation of the rupture disk is not impeded through the installation of the device Take precautions to prevent damage and danger to persons if the rupture disk is actuated Observe information on the rupture disk sticker The position of the rupture disk is indicated on a sticker beside it The transportation guard must be removed For informat...

Страница 26: ...s generally not required Experience shows that zero point adjustment is advisable only in special cases To achieve maximum measuring accuracy even with low flow rates Under extreme process or operating conditions e g very high process temperatures or very high viscosity fluids 6 2 Mounting the measuring device 6 2 1 Required tools For sensor For flanges and other process connections Corresponding ...

Страница 27: ...eck Is the device undamaged visual inspection Does the measuring device conform to the measuring point specifications For example Process temperature 106 Process pressure refer to the section on Pressure temperature ratings in the Technical Information document Ambient temperature Measuring range Has the correct orientation for the sensor been selected According to sensor type According to medium ...

Страница 28: ...ng stranded cables crimper for wire end ferrule 7 1 2 Requirements for connecting cable The connecting cables provided by the customer must fulfill the following requirements Electrical safety In accordance with applicable federal national regulations Permitted temperature range The installation guidelines that apply in the country of installation must be observed The cables must be suitable for t...

Страница 29: ...ational reliability of the measuring device The maximum cable length for individual wire cross sections is specified in the table below Observe the maximum capacitance and inductance per unit length of the cable and the connection values in the Ex documentation Wire cross section Maximum cable length mm2 AWG m ft 0 5 20 70 230 0 75 18 100 328 1 0 17 100 328 1 5 16 200 656 2 5 14 300 984 Cable diam...

Страница 30: ...plug M12x1 thread M20 Options A B C Device plugs 32 Device plugs 32 Option Q 2 x plug M12x1 Order code for Housing Option A compact coated aluminum Option B compact hygienic stainless Option C ultra compact hygienic stainless L L 26 27 A B 1 2 _ 1 2 A0019528 8 Modbus RS485 terminal assignment connection version for use in non hazardous areas and Zone 2 Div 2 1 Power supply DC 24 V 2 Modbus RS485 O...

Страница 31: ...s 32 Option I plug M12x1 Order code for Housing Option A compact coated aluminum Option B compact hygienic stainless Option C ultra compact hygienic stainless L L 62 72 A B 10 20 _ 1 2 A0030219 9 Modbus RS485 terminal assignment connection version for use in intrinsically safe areas connection via Safety Barrier Promass 100 1 Intrinsically safe power supply 2 Modbus RS485 Order code Output 10 L 20...

Страница 32: ...Intrinsically safe area 7 1 4 Pin assignment device plug Supply voltage Promass 100 Device plug for signal transmission with supply voltage device side MODBUS RS485 intrinsically safe 1 2 4 3 5 A0016809 Pin Assignment 1 L Supply voltage intrinsically safe 2 A Modbus RS485 intrinsically safe 3 B 4 L Supply voltage intrinsically safe 5 Grounding shielding Coding Plug socket A Plug Device plug for su...

Страница 33: ...Pay attention to the protection of persons 4 Comply with national installation regulations and guidelines 5 Observe cable specifications 6 Keep the stripped and twisted lengths of cable shield to the ground terminal as short as possible 7 Shield cables fully Grounding of the cable shield NOTICE In systems without potential matching the multiple grounding of the cable shield causes mains frequency ...

Страница 34: ...e federal national installation codes and regulations Comply with local workplace safety regulations Always connect the protective ground cable before connecting additional cables For use in potentially explosive atmospheres observe the information in the device specific Ex documentation The power unit must be tested to ensure it meets safety requirements e g PELV SELV 7 2 1 Connecting the transmi...

Страница 35: ...with the terminal assignment or the device plug pin assignment 7 2 2 Connecting the Safety Barrier Promass 100 In the case of the device version with Modbus RS485 intrinsically safe the transmitter must be connected to the Safety Barrier Promass 100 1 Strip the cable ends In the case of stranded cables also fit ferrules 2 Connect the cable in accordance with the terminal assignment 29 3 Where appl...

Страница 36: ...ifications 28 3 Safety Barrier Promass 100 terminal assignment 32 4 Observe cable specifications 28 5 Non hazardous area 6 Non hazardous area and Zone 2 Div 2 7 Intrinsically safe area 8 Transmitter terminal assignment 29 7 2 3 Ensure potential equalization Requirements No special measures for potential equalization are required For devices intended for use in hazardous locations please observe th...

Страница 37: ...e shield the cable shield must be grounded at both ends to comply with EMC requirements observe cable specifications 28 3 Distribution box 4 Transmitter Modbus RS485 intrinsically safe 2 1 A B 3 A B L L L L A B 5 6 7 4 8 A B L L A0028766 15 Connection example for Modbus RS485 intrinsically safe 1 Control system e g PLC 2 Cable shield observe cable specifications 3 Safety Barrier Promass 100 4 Obse...

Страница 38: ...e device undamaged visual inspection Do the cables used meet the requirements 28 Do the cables have adequate strain relief Are all the cable glands installed firmly tightened and leak tight Cable run with water trap 38 Depending on the device version are all the device plugs firmly tightened Does the supply voltage match the specifications on the transmitter nameplate 98 For device version with Mo...

Страница 39: ... Modbus RS485 Operation options 39 8 Operation options 8 1 Overview of operating options 1 2 A0017760 1 Computer with FieldCare or DeviceCare operating tool via Commubox FXA291 and service interface 2 Control system e g PLC ...

Страница 40: ...ss F 100 Modbus RS485 40 8 2 Structure and function of the operating menu 8 2 1 Structure of the operating menu For an overview of the operating menu for experts Description of Device Parameters document supplied with the device 116 ...

Страница 41: ...er 1 Setup Submenu 1 Submenu n Device tag Advanced setup Enter access code Parameter 1 Parameter n Submenu 1 Submenu n Diagnostics Parameter 1 Parameter n Submenu 1 Submenu n Operating menu for operators and maintenances Parameter n Operator Maintenance Task oriented Function orien ted Expert Wizard 1 Parameter 1 Wizard n Parameter n Parameter n Input A0018237 EN 16 Schematic structure of the oper...

Страница 42: ...all parameters for error detection and analyzing process and device errors Diagnostic list Contains up to 5 currently pending diagnostic messages Event logbook Contains event messages that have occurred Device information Contains information for identifying the device Measured values Contains all current measured values Heartbeat The functionality of the device is checked on demand and the verifi...

Страница 43: ...485 1 2 3 A0030216 1 Service interface CDI of measuring device 2 Commubox FXA291 3 Computer with FieldCare operating tool with COM DTM CDI Communication FXA291 Via service interface CDI E 1 2 3 A0014019 1 Service interface CDI Endress Hauser Common Data Interface of the measuring device 2 Commubox FXA291 3 Computer with FieldCare operating tool with COM DTM CDI Communication FXA291 ...

Страница 44: ...alization of the measured value memory line recorder and event logbook For additional information about FieldCare see Operating Instructions BA00027S and BA00059S Source for device description files See information 46 Establishing a connection 1 Start FieldCare and launch the project 2 In the network Add a device The Add device window opens 3 Select the CDI Communication FXA291 option from the lis...

Страница 45: ...ce name 4 Tag name 5 Status area with status signal 78 6 Display area for current measured values 7 Edit toolbar with additional functions such as save restore event list and create documentation 8 Navigation area with operating menu structure 9 Working area 10 Range of action 11 Status area 8 3 3 DeviceCare Function scope Tool to connect and configure Endress Hauser field devices The fastest way ...

Страница 46: ...re versions for the device 9 1 2 Operating tools The suitable device description file for the individual operating tools is listed in the table below along with information on where the file can be acquired FieldCare www endress com Download Area CD ROM contact Endress Hauser DVD contact Endress Hauser DeviceCare www endress com Download Area CD ROM contact Endress Hauser DVD contact Endress Hause...

Страница 47: ...gister of the measuring device Use function code 16 to write multiple registers with just 1 telegram Write only 1 device parameter Example reset totalizer 08 Diagnostics Master checks the communication connection to the measuring device The following Diagnostics codes are supported Sub function 00 Return query data loopback test Sub function 02 Return diagnostics register 16 Write multiple registe...

Страница 48: ...byte MSB Least significant byte LSB STRING Data length depends on the device parameter e g presentation of a device parameter with a data length 18 bytes 9 registers Byte 17 Byte 16 Byte 1 Byte 0 Most significant byte MSB Least significant byte LSB 9 2 5 Byte transmission sequence Byte addressing i e the transmission sequence of the bytes is not specified in the Modbus specification For this reaso...

Страница 49: ...a map for a maximum of 16 device parameters to allow users to call up multiple device parameters via Modbus RS485 and not only individual device parameters or a group of consecutive device parameters Grouping of device parameters is flexible and the Modbus master can read or write to the entire data block simultaneously with a single request telegram Structure of the Modbus data map The Modbus dat...

Страница 50: ...ia Modbus RS485 Carried out using register addresses 5001 5016 Scan list No Modbus RS485 register Data type Configuration register 0 5001 Integer Scan list register 0 Integer 15 5016 Integer Scan list register 15 Reading out data via Modbus RS485 The Modbus master accesses the data area of the Modbus data map to read out the current values of the device parameters defined in the scan list Master a...

Страница 51: ...Float only Value of scan list register Value of scan list register 15 5081 5082 Integer float Read write Data type depends on the device parameters entered in the scan list Data access depends on the device parameters entered in the scan list If the device parameter entered supports read and write access the parameter can also be accessed via the data area ...

Страница 52: ...re connection For connecting via FieldCare 44 For the FieldCare 45 user interface 10 3 Setting the operating language Factory setting English or ordered local language The operating language can be set in FieldCare or DeviceCare Operation Display language 10 4 Configuring the measuring device The Setup menu with its submenus contains all the parameters needed for standard operation Setup Device ta...

Страница 53: ...ce tag Enter the name for the measuring point Max 32 characters such as letters numbers or special characters e g Promass 10 4 2 Setting the system units In the System units submenu the units of all the measured values can be set Depending on the device version not all submenus and parameters are available in every device The selection can vary depending on the order code Navigation Setup menu Adv...

Страница 54: ...ess variable Unit choose list Country specific l h gal min us Volume unit Select volume unit Unit choose list Country specific l DN 150 6 m gal us Corrected volume flow unit Select corrected volume flow unit Result The selected unit applies for Corrected volume flow parameter 71 Unit choose list Country specific Nl h Sft min Corrected volume unit Select corrected volume unit Unit choose list Count...

Страница 55: ...ernal temperature parameter 6080 Maximum value parameter 6108 Minimum value parameter 6109 Carrier pipe temperature parameter 6027 Maximum value parameter 6029 Minimum value parameter 6030 Reference temperature parameter 1816 Temperature parameter Unit choose list Country specific C F Pressure unit Select process pressure unit Result The unit is taken from Pressure value parameter 57 External pres...

Страница 56: ...d submenu contains parameters that must be configured in order to select and set the medium Navigation Setup menu Medium selection Medium selection Select medium 57 Select gas type 57 Reference sound velocity 57 Temperature coefficient sound velocity 57 Pressure compensation 57 Pressure value 57 External pressure 57 ...

Страница 57: ...e Others option is selected Enter sound velocity of gas at 0 C 32 F 1 to 99999 9999 m s 415 0 m s Temperature coefficient sound velocity The Others option is selected in the Select gas type parameter Enter temperature coefficient for the gas sound velocity Positive floating point number 0 m s K Pressure compensation Select pressure compensation type Off Fixed value External value Off Pressure valu...

Страница 58: ... address Enter device address 1 to 247 247 Baudrate Define data transfer speed 1200 BAUD 2400 BAUD 4800 BAUD 9600 BAUD 19200 BAUD 38400 BAUD 57600 BAUD 115200 BAUD 19200 BAUD Data transfer mode Select data transfer mode ASCII RTU RTU Parity Select parity bits Picklist ASCII option 0 Even option 1 Odd option Picklist RTU option 0 Even option 1 Odd option 2 None 1 stop bit option 3 None 2 stop bits ...

Страница 59: ...tory setting Assign diagnostic behavior Select diagnostic behavior for MODBUS communication Off Alarm or warning Warning Alarm Alarm Failure mode Select measured value output behavior when a diagnostic message occurs via Modbus communication NaN 1 NaN value Last valid value NaN value 1 Not a Number ...

Страница 60: ...ass flow Volume flow Corrected volume flow Mass flow On value low flow cutoff One of the following options is selected in the Assign process variable parameter 60 Mass flow Volume flow Corrected volume flow Enter on value for low flow cut off Positive floating point number Depends on country and nominal diameter Off value low flow cutoff One of the following options is selected in the Assign proce...

Страница 61: ...or partially filled pipe detection Off Density Reference density Off Low value partial filled pipe detection One of the following options is selected in the Assign process variable parameter 61 Density Reference density Enter lower limit value for deactivating partialy filled pipe detection Signed floating point number 200 High value partial filled pipe detection One of the following options is se...

Страница 62: ...Advanced setup Enter access code 62 Calculated values 62 Sensor adjustment 64 Totalizer 1 to n 65 Viscosity Concentration Heartbeat setup Administration 66 10 5 1 Using the parameter to enter the access code Navigation Setup menu Advanced setup Parameter overview with brief description Parameter Description User entry Enter access code Enter access code to disable write protection of parameters 0 ...

Страница 63: ...d reference density Calculated reference density Reference density by API table 53 External reference density Calculated reference density External reference density In the Corrected volume flow calculation parameter the External reference density option is selected Shows external reference density Floating point number with sign Fixed reference density The Fixed reference density option is select...

Страница 64: ...sor adjustment submenu contains parameters that pertain to the functionality of the sensor Navigation Setup menu Advanced setup Sensor adjustment Sensor adjustment Installation direction 64 Zero point adjustment 64 Parameter overview with brief description Parameter Description Selection Factory setting Installation direction Set sign of flow direction to match the direction of the arrow on the se...

Страница 65: ...ction User interface Factory setting Zero point adjustment control Start zero point adjustment Cancel Busy Zero point adjust failure Start Cancel Progress Shows the progress of the process 0 to 100 10 5 4 Configuring the totalizer In the Totalizer 1 to n submenu the individual totalizer can be configured Navigation Setup menu Advanced setup Totalizer 1 to n Totalizer 1 to n Assign process variable...

Страница 66: ...he Assign process variable parameter 66 of the Totalizer 1 to n submenu Volume flow Mass flow Corrected volume flow Target mass flow Carrier mass flow Select totalizer calculation mode Net flow total Forward flow total Reverse flow total Net flow total Failure mode One of the following options is selected in the Assign process variable parameter 66 of the Totalizer 1 to n submenu Volume flow Mass ...

Страница 67: ...e Cancel To delivery settings Restart device Cancel 10 6 Simulation The Simulation submenu enables you to simulate without a real flow situation various process variables in the process and the device alarm mode and to verify downstream signal chains switching valves or closed control loops Navigation Diagnostics menu Simulation Simulation Assign simulation process variable 68 Value process variab...

Страница 68: ...ble selected 0 Simulation device alarm Switch the device alarm on and off Off On Off Visibility depends on order options or device settings 10 7 Protecting settings from unauthorized access The following options exist for protecting the configuration of the measuring device from unintentional modification after commissioning Write protection via write protection switch 68 10 7 1 Write protection v...

Страница 69: ...e protection Setting the write protection switch on the main electronics module to the Off position factory setting disables hardware write protection If hardware write protection is enabled the Locking status parameter displays the Hardware locked option if disabled the Locking status parameter does not display any option 4 Reverse the removal procedure to reassemble the transmitter ...

Страница 70: ...ternal processes running in the device e g data upload download reset etc Once the internal processing has been completed the parameters can be changed once again 11 2 Adjusting the operating language Detailed information To configure the operating language 52 For information on the operating languages supported by the measuring device 113 11 3 Reading measured values With the Measured values subm...

Страница 71: ...splays the mass flow currently measured Dependency The unit is taken from the Mass flow unit parameter 54 Signed floating point number Volume flow Displays the volume flow currently calculated Dependency The unit is taken from the Volume flow unit parameter 54 Signed floating point number Corrected volume flow Displays the corrected volume flow currently calculated Dependency The unit is taken fro...

Страница 72: ...ation unit parameter Signed floating point number Target mass flow With the following conditions Order code for Application package option ED Concentration The WT option or the User conc option is selected in the Concentration unit parameter The software options currently enabled are displayed in the Software option overview parameter Displays the target fluid mass flow currently measured Dependen...

Страница 73: ...alizer counter value Signed floating point number Totalizer overflow One of the following options is selected in the Assign process variable parameter 66 Totalizer 1 to n submenu Volume flow Mass flow Corrected volume flow Target mass flow Carrier mass flow Displays the current totalizer overflow Integer with sign Visibility depends on order options or device settings 11 4 Adapting the measuring d...

Страница 74: ...Totalize Preset value One of the following options is selected in the Assign process variable parameter 66 of the Totalizer 1 to n submenu Volume flow Mass flow Corrected volume flow Target mass flow Carrier mass flow Specify start value for totalizer Dependency The unit of the selected process variable is specified for the totalizer depending on the selection made in the Assign process variable p...

Страница 75: ...ined start value from the Preset value parameter Reset totalize The totalizer is reset to 0 and the totaling process is restarted Preset totalize The totalizer is set to the defined start value from the Preset value parameter and the totaling process is restarted 11 5 2 Function scope of the Reset all totalizers parameter Options Description Cancel No action is executed and the user exits the para...

Страница 76: ... operated outside the application 1 Check and correct parameter configuration 2 Observe limit values specified in the Technical Data For access Error Possible causes Solution No write access to parameters Hardware write protection enabled Set the write protection switch on main electronics module to the OFF position 68 No connection via Modbus RS485 Modbus RS485 bus cable connected incorrectly Che...

Страница 77: ...ight emitting diodes 12 2 1 Transmitter Different LEDs in the transmitter provide information on the device status LED Color Meaning Supply voltage Off Supply voltage is off or too low Green Supply voltage is ok Alarm Off Device status is ok Flashing red A device error of diagnostic behavior Warning has occurred Red A device error of diagnostic behavior Alarm has occurred Boot loader is active Com...

Страница 78: ...red can be shown in the Diagnostics menu Via parameter Via submenu 83 Status signals The status signals provide information on the state and reliability of the device by categorizing the cause of the diagnostic information diagnostic event Symbol Meaning Failure A device error has occurred The measured value is no longer valid Function check The device is in service mode e g during a simulation Ou...

Страница 79: ... the Diagnostics menu 1 Call up the desired parameter 2 On the right in the working area mouse over the parameter A tool tip with remedy information for the diagnostic event appears 12 4 Diagnostic information via communication interface 12 4 1 Reading out diagnostic information Diagnostic information can be read out via Modbus RS485 register addresses Via register address 6821 data type string di...

Страница 80: ...ription Alarm The device stops measurement The measured value output via Modbus RS485 and the totalizers assume the defined alarm condition A diagnostic message is generated Warning The device continues to measure The measured value output via Modbus RS485 and the totalizers are not affected A diagnostic message is generated Logbook entry only The device continues to measure The diagnostic message...

Страница 81: ... 242 Software incompatible 1 Check software 2 Flash or change main electronics module F Alarm 270 Main electronic failure Change main electronic module F Alarm 271 Main electronic failure 1 Restart device 2 Change main electronic module F Alarm 272 Main electronic failure 1 Restart device 2 Contact service F Alarm 273 Main electronic failure Change electronic F Alarm 274 Main electronic failure Ch...

Страница 82: ...ient temperature S Warning 1 834 Process temperature too high Reduce process temperature S Warning 1 835 Process temperature too low Increase process temperature S Warning 1 843 Process limit Check process conditions S Warning 862 Partly filled pipe 1 Check for gas in process 2 Adjust detection limits S Warning 910 Tubes not oscillating 1 Check electronic 2 Inspect sensor F Alarm 912 Medium inhomo...

Страница 83: ...iagnostic information If two or more messages occur simultaneously the message with the highest priority is shown on the display Symbol for diagnostic behavior diagnostic code and short message Previous diagnostics Two diagnostic events have already occurred Shows the diagnostic event that occurred prior to the current diagnostic event along with its diagnostic information Symbol for diagnostic be...

Страница 84: ... includes entries for Diagnostic events 80 Information events 85 In addition to the operation time of its occurrence and possible troubleshooting measures each event is also assigned a symbol that indicates whether the event has occurred or has ended Diagnostic event Occurrence of the event End of the event Information event Occurrence of the event To call up the measures to rectify a diagnostic e...

Страница 85: ...46 Device verification active I1447 Record application reference data I1448 Application reference data recorded I1449 Recording application ref data failed I1450 Monitoring off I1451 Monitoring on I1457 Failed Measured error verification I1459 Failed I O module verification I1460 Failed Sensor integrity verification I1461 Failed Sensor verification I1462 Failed Sensor electronic module verific 12 ...

Страница 86: ...nchanged 12 11 Device information The Device information submenu contains all parameters that display different information for device identification Navigation Diagnostics menu Device information Device information Device tag 86 Serial number 86 Firmware version 87 Device name 87 Order code 87 Extended order code 1 87 Extended order code 2 87 Extended order code 3 87 ENP version 87 Parameter over...

Страница 87: ...composed of letters numbers and certain punctuation marks e g Extended order code 1 Shows the 1st part of the extended order code The extended order code can also be found on the nameplate of the sensor and transmitter in the Ext ord cd field Character string Extended order code 2 Shows the 2nd part of the extended order code The extended order code can also be found on the nameplate of the sensor...

Страница 88: ...mping upper limit value Operating Instructions BA01659O 06 EN 03 14 It is possible to flash the firmware to the current version or the previous version using the service interface For the compatibility of the firmware version with the previous version the installed device description files and operating tools observe the information about the device in the Manufacturer s information document The m...

Страница 89: ...ed materials are adequately resistant Observe the maximum permitted medium temperature for the measuring device 106 13 2 Measuring and test equipment Endress Hauser offers a wide variety of measuring and test equipment such as W M or device tests Your Endress Hauser Sales Center can provide detailed information on the services List of some of the measuring and testing equipment 92 13 3 Endress Hau...

Страница 90: ...e cycle management database 14 2 Spare parts W M Device Viewer www endress com deviceviewer All the spare parts for the measuring device along with the order code are listed here and can be ordered If available users can also download the associated Installation Instructions Measuring device serial number Is located on the nameplate of the device Can be read out via the Serial number parameter 86 ...

Страница 91: ...ice and Connecting the measuring device sections in reverse order Observe the safety instructions 14 5 2 Disposing of the measuring device LWARNING Danger to personnel and environment from fluids that are hazardous to health Ensure that the measuring device and all cavities are free of fluid residues that are hazardous to health or the environment e g substances that have permeated into crevices o...

Страница 92: ...specific accessories Accessories Description Applicator Software for selecting and sizing Endress Hauser measuring devices Choice of measuring devices for industrial requirements Calculation of all the necessary data for identifying the optimum flowmeter e g nominal diameter pressure loss flow velocity and accuracy Graphic illustration of the calculation results Determination of the partial order ...

Страница 93: ...r Common Data Interface and the USB port of a computer or laptop For details see Technical Information TI00405C 15 4 System components Accessories Description Memograph M graphic data manager The Memograph M graphic data manager provides information on all the relevant measured variables Measured values are recorded correctly limit values are monitored and measuring points analyzed The data are st...

Страница 94: ...for its service life use the measuring device only for media against which the process wetted materials are sufficiently resistant 16 2 Function and system design Measuring principle Mass flow measurement based on the Coriolis measuring principle Measuring system The device consists of a transmitter and a sensor The Safety Barrier Promass 100 is part of the scope of supply and must be implemented ...

Страница 95: ...18000 0 to 661 5 40 1 0 to 45000 0 to 1654 50 2 0 to 70000 0 to 2573 80 3 0 to 180000 0 to 6615 100 4 0 to 350000 0 to 12860 150 6 0 to 800000 0 to 29400 250 10 0 to 2200000 0 to 80850 Measuring ranges for gases The full scale values depend on the density of the gas and can be calculated with the formula below max G max F ρG x max G Maximum full scale value for gas kg h max F Maximum full scale va...

Страница 96: ...ge Over 1000 1 Flow rates above the preset full scale value do not override the electronics unit with the result that the totalizer values are registered correctly 16 4 Output Output signal Modbus RS485 Physical interface In accordance with EIA TIA 485 A standard Terminating resistor For device version used in non hazardous areas or Zone 2 Div 2 integrated and can be activated via DIP switches on ...

Страница 97: ...alue 0 Hz Defined value 0 to 12500 Hz Switch output Failure mode Choose from Current status Open Closed Modbus RS485 Failure mode Choose from NaN value instead of current value Last valid value Interface protocol Via digital communication Modbus RS485 Via service interface CDI RJ45 service interface Plain text display With information on cause and remedial measures Light emitting diodes LED Status...

Страница 98: ...sages Supported by the following function codes 06 Write single registers 16 Write multiple registers 23 Read write multiple registers Supported baud rate 1200 BAUD 2400 BAUD 4800 BAUD 9600 BAUD 19200 BAUD 38400 BAUD 57600 BAUD 115200 BAUD Data transfer mode ASCII RTU Data access Each device parameter can be accessed via Modbus RS485 For Modbus register information see Description of device parame...

Страница 99: ...n non hazardous areas and Zone 2 Div 2 90 mA 10 A 0 8 ms Option M Modbus RS485 for use in intrinsically safe areas 145 mA 16 A 0 4 ms Promass 100 safety barrier Order code for Output Maximum Current consumption Maximum switch on current Option M Modbus RS485 for use in intrinsically safe areas 230 mA 10 A 0 8 ms Power supply failure Totalizers stop at the last value measured Depending on the devic...

Страница 100: ...se the Applicator sizing tool 92 Maximum measured error o r of reading 1 g cm 1 kg l T medium temperature Base accuracy Design fundamentals 104 Mass flow and volume flow liquids 0 05 o r PremiumCal order code for Calibration flow option D for mass flow 0 10 o r Mass flow gases 0 35 o r Density liquids Under reference operating conditions Standard density calibration 1 Wide range Density specificat...

Страница 101: ... 40 20 4 15 6500 650 325 130 65 13 25 18000 1800 900 360 180 36 40 45000 4500 2250 900 450 90 50 70000 7000 3500 1400 700 140 80 180000 18000 9000 3600 1800 360 100 350000 35000 17500 7000 3500 700 150 800000 80000 40000 16000 8000 1600 250 2200000 220000 110000 44000 22000 4400 US units DN 1 1 1 10 1 20 1 50 1 100 1 500 inch lb min lb min lb min lb min lb min lb min 73 50 7 350 3 675 1 470 0 735 ...

Страница 102: ...ndamentals 104 Mass flow and volume flow liquids 0 025 o r PremiumCal for mass flow 0 05 o r Mass flow gases 0 25 o r Density liquids 0 00025 g cm3 Temperature 0 25 C 0 0025 T C 0 45 F 0 0015 T 32 F Response time The response time depends on the configuration damping Influence of ambient temperature Current output o r of reading Temperature coefficient Max 0 005 o r C Pulse frequency output Temper...

Страница 103: ...m3 C 0 000025 g cm3 F kg m 3 C 0 0 8 80 160 240 F 0 5 0 100 200 300 0 2 4 6 8 10 2 1 12 14 16 18 320 400 480 560 A0034654 1 Field density calibration for example at 20 C 68 F 2 Special density calibration Temperature 0 005 T C 0 005 T 32 F Influence of medium pressure The table below shows the effect on accuracy of mass flow due to a difference between calibration pressure and process pressure o r...

Страница 104: ... function of the flow rate Flow rate Maximum measured error in o r ZeroPoint BaseAccu 100 A0021332 BaseAccu A0021339 ZeroPoint BaseAccu 100 A0021333 ZeroPoint MeasValue 100 A0021334 Calculation of the maximum repeatability as a function of the flow rate Flow rate Maximum repeatability in o r ZeroPoint BaseRepeat 100 A0021335 A0021340 ZeroPoint BaseRepeat 100 A0021336 ZeroPoint MeasValue 100 A00213...

Страница 105: ...smitter and sensor As standard IP66 67 type 4X enclosure With the order code for Sensor options option CM IP69 can also be ordered When housing is open IP20 type 1 enclosure Display module IP20 type 1 enclosure Safety Barrier Promass 100 IP20 Vibration resistance Vibration sinusoidal according to IEC 60068 2 6 2 to 8 4 Hz 3 5 mm peak 8 4 to 2000 Hz 1 g peak Vibration broad band random according to...

Страница 106: ...erature Tm Medium temperature A Maximum permitted medium temperature Tm at Ta max 60 C 140 F higher medium temperatures Tm require a reduced ambient temperature Ta B Maximum permitted ambient temperature Ta for the maximum specified medium temperature Tm of the sensor Values for devices used in the hazardous area Separate Ex documentation XA for the device Not insulated Insulated A B A B Version T...

Страница 107: ...er pressure is decisive for the maximum nominal pressure 108 The secondary containment burst pressure refers to a typical internal pressure achieved prior to mechanical failure of the secondary containment as determined during type testing The corresponding type test declaration can be ordered with the device order code for Additional approval option LN Type test containment DN Secondary containme...

Страница 108: ...A rupture disk Rupture disks cannot be combined with the separately available heating jacket Special mounting instructions 25 For information on the dimensions see the Mechanical construction section Flow limit Select the nominal diameter by optimizing between the required flow range and permissible pressure loss For an overview of the full scale values for the measuring range see the Measuring ra...

Страница 109: ...ght exclusive of packaging material refer to devices with EN DIN PN 40 flanges Weight specifications including transmitter order code for Housing option A Compact aluminum coated Different values due to different transmitter versions Weight in SI units DN mm Weight kg 8 9 15 10 25 12 40 17 50 28 80 53 100 94 150 152 250 398 Weight in US units DN in Weight lbs 3 8 20 22 1 26 1 37 2 62 3 117 4 207 6...

Страница 110: ...steel 1 4404 316L Cable entries cable glands 1 2 3 A0020640 18 Possible cable entries cable glands 1 Female thread M20 1 5 2 Cable gland M20 1 5 3 Adapter for cable entry with internal thread G or NPT Order code for Housing option A Compact coated aluminum The various cable entries are suitable for hazardous and non hazardous areas Cable entry cable gland Material Cable gland M20 1 5 Nickel plated...

Страница 111: ...tube material option HA TH Alloy C22 Stainless steel 1 4404 316L Measuring tubes DN 8 to 100 3 8 4 stainless steel 1 4539 904L Manifold stainless steel 1 4404 316 316L DN 150 6 DN 250 10 stainless steel 1 4404 316 316L Manifold stainless steel 1 4404 316 316L DN 8 to 250 3 8 to 10 Alloy C22 2 4602 UNS N06022 Manifold Alloy C22 2 4602 UNS N06022 Process connections Flanges according to EN 1092 1 DI...

Страница 112: ... 1145 thread ISO 2853 thread ISO 2037 DIN 11864 1 Form A thread DIN 11866 series A VCO connections 8 VCO 4 12 VCO 4 For information on the different materials used in the process connections Surface roughness All data relate to parts in contact with fluid The following surface roughness quality can be ordered Not polished Ramax 0 8 µm 32 µin Ramax 0 4 µm 16 µin 16 11 Operability Service interface ...

Страница 113: ...EHEDG To meet the requirements for EHEDG certification the device must be used with process connections in accordance with the EHEDG position paper entitled Easy Cleanable Pipe Couplings and Process Connections www ehedg org Modbus RS485 certification The measuring device meets all the requirements of the MODBUS TCP conformity test and has the MODBUS TCP Conformance Test Policy Version 2 0 The mea...

Страница 114: ...Specifications for integrating fieldbus devices in engineering tools for field devices NAMUR NE 107 Self monitoring and diagnosis of field devices NAMUR NE 131 Requirements for field devices for standard applications NAMUR NE 132 Coriolis mass meter NACE MR0103 Materials resistant to sulfide stress cracking in corrosive petroleum refining environments NACE MR0175 ISO 15156 1 Materials for use in H...

Страница 115: ...on Package Description Concentration measurement and special density Calculation and outputting of fluid concentrations Many applications use density as a key measured value for monitoring quality or controlling processes The device measures the density of the fluid as standard and makes this value available to the control system The Special Density application package offers high precision densit...

Страница 116: ...e Promass F 100 TI01318O Description of Device Parameters Measuring device Documentation code Proline Promass 100 GP01103O Supplementary device dependent documentation Safety Instructions Content Documentation code ATEX IECEx Ex i XA00159D ATEX IECEx Ex nA XA01029D cCSAus IS XA00160D INMETRO Ex i XA01219D INMETRO Ex nA XA01220D NEPSI Ex i XA01249D NEPSI Ex nA XA01262D Special Documentation Content...

Страница 117: ...7 Installation Instructions Contents Comment Installation instructions for spare part sets and accessories Access the overview of all the available spare part sets via W M Device Viewer 90 Accessories available for order with Installation Instructions 92 ...

Страница 118: ...tion 99 D Declaration of Conformity 12 Degree of protection 38 105 Density 106 Design Measuring device 13 Design fundamentals Maximum measured error 104 Repeatability 104 Designated use 10 Device components 13 Device description files 46 Device documentation Supplementary documentation 8 Device locking status 70 Device name Sensor 16 Transmitter 15 Device repair 90 Device revision 46 Device type I...

Страница 119: ...pressure 103 Medium temperature 102 Information on the document 6 Inlet runs 22 Input 95 Inspection Received goods 14 Inspection check Connection 38 Installation 20 Installation conditions Down pipe 20 Inlet and outlet runs 22 Installation dimensions 22 Mounting location 20 Orientation 21 Rupture disk 25 Sensor heating 24 System pressure 23 Thermal insulation 23 Vibrations 25 Installation dimensio...

Страница 120: ...formation Submenu 86 Diagnostics Menu 83 Low flow cut off Wizard 60 Measured variables Submenu 70 Medium selection Submenu 56 Partially filled pipe detection Wizard 61 Sensor adjustment Submenu 64 Setup Menu 53 Simulation Submenu 67 System units Submenu 53 Totalizer Submenu 72 Totalizer 1 to n Submenu 65 Totalizer handling Submenu 73 Zero point adjustment Submenu 64 Performance characteristics 100...

Страница 121: ...erview 42 Process variables 62 Sensor adjustment 64 Simulation 67 System units 53 Totalizer 72 Totalizer 1 to n 65 Totalizer handling 73 Zero point adjustment 64 Supply voltage 98 Surface roughness 112 System design Measuring system 94 see Measuring device design System integration 46 System pressure 23 T Technical data overview 94 Temperature range Medium temperature 106 Storage temperature 18 Te...

Страница 122: ...e 1 25474 Ellerbek Germany P 49 4101 304 0 The specifications contained herein are subject to change without notice and any user of said specifications should verify from the manufacturer that the specifications are currently in ef fect Otherwise the manufacturer assumes no responsibility for the use of specifications which may have been changed and are no longer in effect Contact information is subje...