Proline Promass E 100 Modbus RS485
Technical data
Hauser
85
Interface/protocol
• Via digital communication:
Modbus RS485
• Via service interface
CDI-RJ45 service interface
Plain text display
With information on cause and remedial measures
Light emitting diodes (LED)
Status information
Status indicated by various light emitting diodes
The following information is displayed depending on the device version:
• Supply voltage active
• Data transmission active
• Device alarm/error has occurred
Diagnostic information via light emitting diodes → 67
Low flow cut off
The switch points for low flow cut off are user-selectable.
Galvanic isolation
The following connections are galvanically isolated from each other:
• Outputs
• Power supply
Protocol-specific data
Protocol-specific data
Protocol
Modbus Applications Protocol Specification V1.1
Device type
Slave
Slave address range
1 to 247
Broadcast address range
0
Function codes
• 03: Read holding register
• 04: Read input register
• 06: Write single registers
• 08: Diagnostics
• 16: Write multiple registers
• 23: Read/write multiple registers
Broadcast messages
Supported by the following function codes:
• 06: Write single registers
• 16: Write multiple registers
• 23: Read/write multiple registers
Supported baud rate
• 1
200 BAUD
• 2
400 BAUD
• 4
800 BAUD
• 9
600 BAUD
• 19
200 BAUD
• 38
400 BAUD
• 57
600 BAUD
• 115
200 BAUD
Data transfer mode
• ASCII
• RTU
Data access
Each device parameter can be accessed via Modbus RS485.
For Modbus register information, see "Description of device parameters"
documentation