![Endress+Hauser Proline Promass E 100 Скачать руководство пользователя страница 67](http://html1.mh-extra.com/html/endress-hauser/proline-promass-e-100/proline-promass-e-100_operating-instructions-manual_2408129067.webp)
Proline Promass E 100 Modbus RS485
Diagnostics and troubleshooting
Hauser
67
12
Diagnostics and troubleshooting
12.1 General troubleshooting
For output signals
Problem
Possible causes
Remedy
Green power LED on the main
electronics module of the
transmitter is dark
Supply voltage does not match that
specified on the nameplate.
Apply the correct supply voltage
→ 30.
Green power LED on the main
electronics module of the
transmitter is dark
Power supply cable connected
incorrectly
Check the terminal assignment .
Green power LED on Safety Barrier
Promass 100 is dark
Supply voltage does not match that
specified on the nameplate.
Apply the correct supply voltage
→ 30.
Green power LED on Safety Barrier
Promass 100 is dark
Power supply cable connected
incorrectly
Check the terminal assignment .
Device measures incorrectly.
Configuration error or device is
operated outside the application.
1. Check and correct parameter
configuration.
2. Observe limit values specified in
the "Technical Data".
For access
Problem
Possible causes
Remedy
No write access to parameters
Hardware write protection enabled
Set the write protection switch on
the main electronics module to the
OFF position → 61.
No connection via Modbus RS485
Modbus RS485 bus cable connected
incorrectly
Check the terminal assignment .
No connection via Modbus RS485
Device plug connected incorrectly
Check the pin assignment of the
device plug .
No connection via Modbus RS485
Modbus RS485 cable incorrectly
terminated
Check terminating resistor .
No connection via Modbus RS485
Incorrect settings for the
communication interface
Check the Modbus RS485
configuration → 53.
No connection via service interface
Incorrect configuration of USB
interface on PC or driver not
installed correctly.
Observe the documentation for the
Commubox.
FXA291: Document
"Technical Information"
TI00405C
12.2 Diagnostic information via light emitting diodes
12.2.1 Transmitter
Various light emitting diodes (LEDs) on the main electronics module of the transmitter
provide information on device status.
LED
Color
Meaning
Power
Off
Supply voltage is off or too low.
Green
Supply voltage is ok.
Содержание Proline Promass E 100