Wiring
Commissioning Instructions for PTB gas approval Proline Promass 84
14
Hauser
3.3
Electrical connection of remote version
a0003681
Fig. 6:
Connecting the remote version
a
Wall-mount housing: Non-hazardous area and ATEX II3G / Zone 2
b
Wall-mount housing: ATEX II2G / Zone 1 /FM/CSA
c
Remote version, flange version
Terminal no.:
4/5
= gray
6/7
= green
8
= yellow
9/10 = pink
11/12 = white
41/42 = brown
3.3.1
Cable specification, connecting cable
In the remote version, the specifications of the cable connecting the transmitter to the sensor are as
follows:
• 6 × 0.38 mm² PVC cable with common shield and individually shielded cores
• Conductor resistance:
50
/km
• Capacitance core/shield:
420 pF/m
• Cable length: max. 20 m
• Operating temperature: max. +105 °C
!
Note!
The cable must be installed securely, to prevent movement.
4
5
6
7
8
9
10
11 12
41 42
4
5
6
7
8
9
10
11 12
41 42
S1 S1 S2 S2 GND TM TM TT TT
+
+
+
+
+
+
+
+
S1 S1 S2 S2 GND TM TM TT TT
a
b
c
d
d
d
e