Proline Promag H 300
82
Hauser
Display and operating elements
The display and operating elements correspond to those of the display module → 81.
• The measuring device is always supplied with a dummy cover when the remote display and
operating module DKX001 is used. Display or operation at the transmitter is not possible in
this case.
• The remote display and operating module DKX001 can also be ordered separately and
subsequently as an accessory without a measuring device → 91.
• If ordered subsequently: The remote display and operating module DKX001 cannot be
connected at the same time as the existing display or operation unit. Only one display or
operation unit may be connected to the transmitter at any one time.
Material
The housing material of the display and operating module DKX001 corresponds to the selected
material of the transmitter housing.
Transmitter housing
Remote display and operating module
Order code for "Housing"
Material
Material
Option
A
"Aluminum, coated"
AlSi10Mg, coated
AlSi10Mg, coated
Cable entry
Corresponds to the choice of transmitter housing, order code for "Electrical connection".
Connecting cable
Dimensions
Remote operation
Via HART protocol
This communication interface is available in device versions with a HART output.
1
2
3
5
7
4
6
A0028747
48
Options for remote operation via HART protocol (active)
1
Control system (e.g. PLC)
2
Field Communicator 475
3
Computer with Web browser (e.g. Internet Explorer) for accessing the integrated device Web server or
computer with operating tool (e.g. FieldCare, DeviceCare, AMS Device Manager, SIMATIC PDM) with COM
DTM "CDI Communication TCP/IP"
4
Commubox FXA195 (USB)
5
Field Xpert SFX350 or SFX370
6
VIATOR Bluetooth modem with connecting cable
7
Transmitter