Proline Promag 50D
10
Hauser
Performance characteristics
Reference operating
conditions
As per DIN EN 29104 and VDI/VDE 2641
• Fluid temperature: +28 °C ±2 K
• Ambient temperature: +22 °C ±2 K
• Warm-up period: 30 minutes
Installation
• Inlet run >10 x DN
• Outlet run > 5 x DN
• Sensor and transmitter grounded.
• The sensor is centered in the pipe.
Maximum measured error
• Current output: also typically ±5
μ
A
• Pulse output: ±0.5% o.r. ±1 mm/s (o.r. = of reading)
Fluctuations in the supply voltage do not have any effect within the specified range.
a0003200
Max. measured error in % of reading
Repeatability
Max. ±0.1% o.r. ±0.5 mm/s (o.r. = of reading)
0
1
2.5
[%]
2.0
1.5
1.0
0.5
0
2
4
6
8
10
v [m/s]
0.5 %