Description of device parameters
Proline Promag 300 Modbus RS485
204
Hauser
Additional information
Description
The desired simulation value is defined in the
Signal level
parameter (→ 204).
Selection
• Off
Simulation for the status input is switched off. The device is in normal measuring mode
or another process variable is being simulated.
• On
Simulation for the status input is active.
Signal level
Navigation
Expert → Diagnostics → Simulation → Signal level (1356)
Prerequisite
In the
Status inp. sim.
On
option is selected.
Description
Use this function to select the signal level for the simulation of the status input. In this
way, users can verify the correct configuration of the status input and the correct function
of upstream feed-in units.
Selection
• High
• Low
Curr.inp 1 to n sim.
Navigation
Expert → Diagnostics → Simulation → Curr.inp 1 to n sim. (1608–1 to n)
Description
Option for switching simulation of the current input on and off. The display alternates
between the measured value and a diagnostic message of the "Function check" category (C)
while simulation is in progress.
The desired simulation value is defined in the
Value curr.inp 1 to n
parameter.
Selection
• Off
• On
Factory setting
Off
Additional information
Selection
• Off
Current simulation is switched off. The device is in normal measuring mode or another
process variable is being simulated.
• On
Current simulation is active.