Proline Promag 300 Modbus RS485
Description of device parameters
Hauser
203
• Flow velocity
• Conductivity
• CorrConductivity
• Temperature
Factory setting
Off
Additional information
Description
The simulation value of the process variable selected is defined in the
Proc. var. value
Proc. var. value
Navigation
Expert → Diagnostics → Simulation → Proc. var. value (1811)
Description
Use this function to enter a simulation value for the selected process variable. Subsequent
measured value processing and the signal output use this simulation value. In this way,
users can verify whether the measuring device has been configured correctly.
User entry
Depends on the process variable selected
Factory setting
0
Additional information
Entry
The unit of the displayed measured value is taken from the
System units
submenu
Status inp. sim.
Navigation
Expert → Diagnostics → Simulation → Status inp. sim. (1355)
Description
Use this function to switch simulation of the status input on and off. The display alternates
between the measured value and a diagnostic message of the "Function check" category (C)
while simulation is in progress.
Selection
• Off
• On
Factory setting
Off
*
Visibility depends on order options or device settings