Description of device parameters
Proline Promag 300 Modbus RS485
122
Hauser
U[V]
t
24
0
A0026885
9
Alarm, low level
Assign pulse 1 to n
Navigation
Expert → Output → PFS output 1 to n → Assign pulse 1 to n (0460–1 to n)
Prerequisite
The
Pulse
option is selected in the
Operating mode
Description
Use this function to select the process variable for the pulse output.
Selection
• Off
• Volume flow
• Mass flow
• Correct.vol.flow
Factory setting
Off
Value per pulse
Navigation
Expert → Output → PFS output 1 to n → Value per pulse (0455–1 to n)
Prerequisite
In the
Operating mode
Pulse
option is selected, and one of
the following options is selected in the
Assign pulse
• Mass flow
• Volume flow
• Correct.vol.flow
Description
Use this function to enter the value for the measured value that a pulse is equivalent to.
User entry
Signed floating-point number
Factory setting
Depends on country and nominal diameter→ 213
Additional information
Entry
Weighting of the pulse output with a quantity.
The lower the pulse value, the
• better the resolution.
• the higher the frequency of the pulse response.