Device functions - vortex measuring devices
Proline Fieldcheck
86
Hauser
Analyzing the results
During verification, the current verification test section is displayed in the 'HOME position'. The
results can be shown on the display once verification is complete. They can be filed in Fieldcheck
under a storage location number of your choice (up to max. 20) together with the measuring device
serial no. and the tag name. Via a connection with the FieldCare software, a report of the results
can later be printed out and the data can be archived.
Function group APPLICATION/OUTPUTS
During verification, the function and accuracy of the current output and the optional pulse output
can be checked. Fieldcheck determines the outputs available in the measuring device.
Function description
FUNCTION
VERIFICATION
APPLICATION/OUTPUTS
FLOW SPECIFICATION
(only Prowirl 73)
Use this function to select the type of flow unit for the verification of the amplifier.
Options:
VOLUME FLOW
MASS FLOW
CORRECTED VOLUME FLOW
!
Note!
The selection of the flow units can be restricted by the configured fluid:
MASS FLOW
not available for gas volume and liquid volume
CORRECTED VOLUME FLOW
not available for saturated steam, superheated steam,
gas volume and liquid volume
OUTPUTS
Use this function to display the outputs determined in the flowmeter on the local display
with the related terminals and to select the outputs to be tested.
1.
Using the
P
keys, select the outputs to be tested during the verification process.
2.
Confirm the option selected with the
F
key.
!
Note!
Start the verification process in the VERIFICATE function
The following combinations are possible for Proline Prowirl:
• I
• I+P
!
Note!
Fieldbus interfaces cannot be tested.