Electrical connection
Flowmeter Proline 500 – digital
24
Hauser
Connecting the connecting cable to the transmitter
1.
3.
2.
4.
4 x
5.
7.
TX 20
TX 20
6.
10 (0.4)
22 mm
8.
24 mm
A0029597
1. Loosen the 4 fixing screws on the housing cover.
2. Open the housing cover.
3. Fold open the terminal cover.
4. Push the cable through the cable entry. To ensure tight sealing, do not remove the
sealing ring from the cable entry.
5. Strip the cable and cable ends. In the case of stranded cables, fit ferrules.
6. Connect the protective ground.
7. Connect the cable according to the terminal assignment for the connecting cable
8. Firmly tighten the cable glands.
This concludes the process for connecting the connecting cable.
9. Close the housing cover.
10. Tighten the securing screw of the housing cover.
11. After connecting the connecting cable:
Connect the signal cable and the supply voltage cable → 25.