Code
Type
Cause and Remedy
E 101-106 Alarm
Fault in instrument electronics
-
Call Hauser Service
E 107
Alarm
Battery voltage too low
-
Make back-up of entered parameters immediately
-
Have battery changed at once by trained personel or ring for service
E 201-202 Alarm
Fault in probe on channel 1 (f < 35 Hz; f > 3000 Hz)
-
Check probe and electronic insert
E 301-302 Alarm
Fault in probe on channel 2 (f < 35 Hz; f > 3000 Hz)
-
Check probe and electronic insert
E 400
Alarm
Fault in probe on channels 1 + 2
-
Check probe, electronic insert and wiring
-
Incorrect mode selected (only one probe)
E 401
Alarm
Fault in probe or wiring, channel 1
-
Check probe, electronic insert and wiring
-
Incorrect operating mode
E 402
Alarm
Fault in probe or wiring, channel 2
-
Check probe, electronic insert and wiring
-
Incorrect operating mode
E 403
Alarm
Difference in measured values between channel 1 and 2 too large
(modes 3 and 5 only)
-
Check calibration and linearisation
E 600
Warning
PFM transmission internal code check
-
can be ignored if it appears only briefly
E 601
Warning
PFM transmission internal code check
-
can be ignored if it appears only briefly
E 602
Warning
Linearisation channel 1does not rise monotonously (volume does not
increase with level)
-
Check and re-enter correct values, reactivate linearization
E 603
Warning
Linearisation channel 2 does not rise monotonously (volume does not
increase with level)
-
Check and re-enter correct values, reactivate linearization
E 604
Warning
Linearisation channel 1 has less than two sets of values
-
Enter more values, reactivate linearization
E 605
Warning
Linearisation channel 2 has less than two sets of values
-
Enter more values, reactivate linearization
E 608
Warning
Value in V0H5 greater than that in V0H6
-
Check input
E 609
Warning
Value in V4H5 greater than that in V4H6
-
Check input
E 610
Warning
Calibration fault, channel 1 (»empty« level > »full« level)
-
Repeat calibration
E 611
Warning
Calibration fault, channel 2 (»empty« level > »full« level)
-
Repeat calibration
E 613
Warning
Instrument in simulation mode, channel 1
-
Switch back when finished
E 614
Warning
Instrument in simulation mode, channel 2
-
Switch back when finished