6
Commissioning
6.1
Function check
Before first commissioning, check if:
• the sensor is correctly installed
• the electrical connection is correct.
If using an assembly with automatic cleaning, check the correct connection of the cleaning agent
(e.g. water or air).
WARNING
!
Escaping process medium
Risk of injury from high pressure, high temperatures or chemical hazards
►
Before applying compressed air to an assembly with cleaning facility, make sure the connections
are correctly fitted.
►
Do not install the assembly in the process if you cannot make the correct connection reliably.
6.2
Calibration
The sensor is calibrated at the factory. A new calibration is only needed in special situations.
6.3
Automatic cleaning
Compressed air is most suitable for cyclic cleaning. The cleaning unit is either ready supplied or can
be retrofitted, and is attached to the sensor head. It operates at a capacity of 20-60 l/min. Optimum
results are achieved with 2 bar (29 psi) and 60 l/min.
The following settings are recommended for the cleaning unit:
Type of soiling
Cleaning interval
Cleaning duration
Media containing grease and oils
15 min
20 s
Biofilm
60 min
20 s