Installation
Oxymax COS61
10
Hauser
3.4
Installation examples
3.4.1
Immersion operation
Upright post and chain assembly
For large basins, where sufficient installation distance is required from the basin edge (aeration
basin, especially), it is advisable to use the upright post and chain assembly. The free swinging of
the immersed assembly practically rules out vibrations from the upright post.
A good self-cleaning of the fluorescence cap is reached due to the swinging of the assembly.
According to this effect, the sensor life time can be extended.
a0013510
Fig. 5:
Chain holder, rail mounted
1
Chain
2
Flexdip CYH112 holder
3
Rail
4
Oxymax sensor
5
Flexdip CYA112 wastewater assembly
a0013511
Fig. 6:
Chain holder, mounted to a post
1
Weather protection cover
2
#
ontroller
/ transmitter
3
Chain
4
Flexdip CYA112 wastewater assembly
5
Oxymax sensor
6
Flexdip CYH112 holder