Maintenance

OUSAF11

18

Hauser

Maintenance checklist

• Replace lamp

The lamp is typically replaced after 8000 to 10 000 operating hours (→   26).

• Replace O-rings in contact with the medium

The replacement of O-rings in contact with the medium depends on the specific

requirements of the process.

Never re-use a used O-ring.

8.2 

Clean sensor

Cleaning the sensor

If the sensor is dirty, this can influence the measurement results and even cause a

malfunction. Therefore the sensor must be cleaned regularly in order to guarantee reliable

measurement results. The frequency and intensity of the cleaning process depend on the

medium. Clean the sensor:

• before every calibration/adjustment/zero point adjustment

• before you send the sensor in for repair

Fouling

Cleaning

Lime deposits

Immerse the sensor in 1-5 % hydrochloric acid (for a few minutes).

Dirt particles on the optical windows

Fold the cloth and wipe the cell.

NOTICE

Cleaning agent residue

Cleaning agent residue can affect the measurement.

Rinse the sensor carefully with water after every cleaning.

8.3 

Replace lamp

Switch off the lamp at the transmitter using the software function.

Allow the lamp to cool down (30 minutes).

1.

Release the cable gland and pull it over the cable to remove it from the sensor.

Содержание OUSAF11

Страница 1: ...Products Solutions Services Operating Instructions OUSAF11 Optical sensor for the measurement of VIS NIR absorption BA00474C 07 EN 16 16 71330142...

Страница 2: ......

Страница 3: ...ct identification 9 4 3 Scope of delivery 10 4 4 Certificates and approvals 10 5 Installation 11 5 1 Installation conditions 11 5 2 Mounting the sensor 14 5 3 Post installation check 15 6 Electrical c...

Страница 4: ...situation Failure to avoid the dangerous situation can result in a fatal or serious injury LCAUTION Causes consequences Consequences of non compliance if applicable Corrective action This symbol alert...

Страница 5: ...OUSAF11 Document information Endress Hauser 5 1 3 Symbols on the product Symbol Meaning Reference to device documentation...

Страница 6: ...lications such as Interphase detection Applications in the food and beverage industry Milk detection in CIP solutions Phase separation e g milk water Product loss detection in outlet Measurement of so...

Страница 7: ...and safeguard them to ensure that they are not operated inadvertently 2 5 Product safety The product is designed to meet state of the art safety requirements has been tested and left the factory in a...

Страница 8: ...ght source A Absorption Extinction coefficient c Concentration OPL Optical path length A light source emits radiation through the medium and the incident radiation is measured on the detector side The...

Страница 9: ...rotected against impact and moisture The original packaging offers the best protection The permitted ambient conditions must be observed see Technical data If you have any questions please contact you...

Страница 10: ...you have any questions please contact your supplier or your local sales center 4 4 Certificates and approvals 4 4 1 mark Declaration of Conformity The product meets the requirements of the harmonized...

Страница 11: ...ns 5 1 1 Measuring system An optical measuring system comprises OUSAF11 sensor photometer Transmitter for example Liquiline CM44P Sensor fixed cable 1 3 2 A0029243 3 Example of a measuring system with...

Страница 12: ...rnal thread 5 1 3 Mounting angles The sensor can be installed up to the horizontal in an assembly support or suitable process connection Other angles of inclination are not recommended Do not install...

Страница 13: ...ion Endress Hauser 13 A A0029248 5 Mounting angles A Permitted mounting angle 0 to 180 5 1 4 Mounting in pipes 4 1 5 2 3 Detail A Detail A 90 5 3 2 A0029259 6 Permitted and unacceptable installation p...

Страница 14: ...m 5 B Lateral view 270 in relation to A arrow indicates the direction of flow C 3D view Do not install the sensor in places where air pockets or bubbles occur 6 13 item 3 or where sedimentation may oc...

Страница 15: ...chnician must have read and understood these Operating Instructions and must follow the instructions contained therein Prior to commencing connection work ensure that no voltage is present on any cabl...

Страница 16: ...es are not fully secured The display is not fully secured risk of moisture entering due to inadequate sealing Loose or insufficiently tightened cables cable ends Conductive cable strands are left in t...

Страница 17: ...ss and process control When carrying out any work on the system take into account possible repercussions for process control or the process itself For your own safety only use genuine accessories With...

Страница 18: ...liable measurement results The frequency and intensity of the cleaning process depend on the medium Clean the sensor before every calibration adjustment zero point adjustment before you send the senso...

Страница 19: ...ctions Only turn the shaft Never turn by the sensor head Make sure that the cable is not twisted Hold the sensor head firmly in position and unscrew the sensor shaft from the sensor head 3 Using the A...

Страница 20: ...nto the sensor head until it is firmly seated in the correct position Tighten the securing screw slightly 8 Mount the sensor again In doing so use PTFE tape to seal the thread of the sensor head Check...

Страница 21: ...sensor no longer functions Only turn the shaft Never turn by the sensor head Make sure that the cable is not twisted Hold the sensor head firmly in position and unscrew the sensor shaft from the senso...

Страница 22: ...e one O ring on either side of the process connection Always replace the two O rings 5 a Release the O ring item a from the guide in the sensor shaft 6 Pull the O ring over the lamp and detector unit...

Страница 23: ...g product was ordered or delivered As an ISO certified company and also due to legal regulations Endress Hauser is obliged to follow certain procedures when handling any returned products that have be...

Страница 24: ...onfigurator on the product page www endress com cyh112 Technical Information TI00430C 11 Technical data 11 1 Input 11 1 1 Measured variable VIS NIR absorption 11 1 2 Measuring range 0 to 3 AU 0 to 6 O...

Страница 25: ...auser 25 11 3 Process 11 3 1 Process temperature 0 to 90 C 32 to 194 F continuous Max 130 C 266 F for 2 hours 11 3 2 Process pressure Max 10 bar 150 psi absolute at 20 C 68 F For version with Tri Clam...

Страница 26: ...ensor shaft Stainless steel 316L O ring EPDM Cable connector ends Nickel plated brass 11 4 4 Process connections Depends on version Thread G1 NPT 1 thread Tri Clamp 2 2 5 3 VARIVENT N DN40 125 11 4 5...

Страница 27: ...6 M Maintenance schedule 17 Measured variable 24 Measuring principle 8 Measuring range 24 Measuring system 11 Mounting angles 12 Mounting the sensor 14 N Nameplate 9 O Occupational safety 6 Operationa...

Страница 28: ...www addresses endress com 71330142 71330142...

Отзывы: