background image

Installation

OUSAF11

14

Hauser

Comply with the following conditions. Otherwise you run the risk of damaging the measuring

point or obtaining incorrect measured values.

The diameter of the pipe must be at least 50 mm (2").

Install the sensor in places with consistent flow conditions.

The best installation location is in the ascending pipe (item 1).

Installation in the horizontal pipe (item 5) is also possible. In this case, you must turn the

sensor so that the measuring slit is aligned with the direction of flow at an angle of at least

12° but not more than 45°. → diagram

12 ... max. 45°

B

A

C

  A0031255

 7

3-A-permitted sensor orientation when installed in horizontal pipes

A

→   13, item 5

B

Lateral view (270° in relation to A), arrow indicates the direction of flow

C

3D view

Do not install the sensor in places where air pockets or bubbles occur (→   6,  13, item

3) or where sedimentation may occur (item 2).

Avoid installation in the down pipe (item 4).

Align the sensor in such a way that the medium flows through the measuring gap (self-

cleaning effect).

5.2 

Mounting the sensor

NOTICE

Mounting errors

Possibility of sensor damage, twisted cables or similar

Make sure that the sensor body is protected against damage from external forces - such as

trolleys on adjacent paths.

Make sure to avoid exerting excessive tensile force on the cable (e.g. from jerky pulling

movements).

Make sure to observe the national grounding regulations when using metal assemblies.

The sensor can be installed directly in tanks and pipes with a suitable process connection, or

in a suitable assembly.

Содержание OUSAF11

Страница 1: ...Products Solutions Services Operating Instructions OUSAF11 Optical sensor for the measurement of VIS NIR absorption BA00474C 07 EN 16 16 71330142...

Страница 2: ......

Страница 3: ...ct identification 9 4 3 Scope of delivery 10 4 4 Certificates and approvals 10 5 Installation 11 5 1 Installation conditions 11 5 2 Mounting the sensor 14 5 3 Post installation check 15 6 Electrical c...

Страница 4: ...situation Failure to avoid the dangerous situation can result in a fatal or serious injury LCAUTION Causes consequences Consequences of non compliance if applicable Corrective action This symbol alert...

Страница 5: ...OUSAF11 Document information Endress Hauser 5 1 3 Symbols on the product Symbol Meaning Reference to device documentation...

Страница 6: ...lications such as Interphase detection Applications in the food and beverage industry Milk detection in CIP solutions Phase separation e g milk water Product loss detection in outlet Measurement of so...

Страница 7: ...and safeguard them to ensure that they are not operated inadvertently 2 5 Product safety The product is designed to meet state of the art safety requirements has been tested and left the factory in a...

Страница 8: ...ght source A Absorption Extinction coefficient c Concentration OPL Optical path length A light source emits radiation through the medium and the incident radiation is measured on the detector side The...

Страница 9: ...rotected against impact and moisture The original packaging offers the best protection The permitted ambient conditions must be observed see Technical data If you have any questions please contact you...

Страница 10: ...you have any questions please contact your supplier or your local sales center 4 4 Certificates and approvals 4 4 1 mark Declaration of Conformity The product meets the requirements of the harmonized...

Страница 11: ...ns 5 1 1 Measuring system An optical measuring system comprises OUSAF11 sensor photometer Transmitter for example Liquiline CM44P Sensor fixed cable 1 3 2 A0029243 3 Example of a measuring system with...

Страница 12: ...rnal thread 5 1 3 Mounting angles The sensor can be installed up to the horizontal in an assembly support or suitable process connection Other angles of inclination are not recommended Do not install...

Страница 13: ...ion Endress Hauser 13 A A0029248 5 Mounting angles A Permitted mounting angle 0 to 180 5 1 4 Mounting in pipes 4 1 5 2 3 Detail A Detail A 90 5 3 2 A0029259 6 Permitted and unacceptable installation p...

Страница 14: ...m 5 B Lateral view 270 in relation to A arrow indicates the direction of flow C 3D view Do not install the sensor in places where air pockets or bubbles occur 6 13 item 3 or where sedimentation may oc...

Страница 15: ...chnician must have read and understood these Operating Instructions and must follow the instructions contained therein Prior to commencing connection work ensure that no voltage is present on any cabl...

Страница 16: ...es are not fully secured The display is not fully secured risk of moisture entering due to inadequate sealing Loose or insufficiently tightened cables cable ends Conductive cable strands are left in t...

Страница 17: ...ss and process control When carrying out any work on the system take into account possible repercussions for process control or the process itself For your own safety only use genuine accessories With...

Страница 18: ...liable measurement results The frequency and intensity of the cleaning process depend on the medium Clean the sensor before every calibration adjustment zero point adjustment before you send the senso...

Страница 19: ...ctions Only turn the shaft Never turn by the sensor head Make sure that the cable is not twisted Hold the sensor head firmly in position and unscrew the sensor shaft from the sensor head 3 Using the A...

Страница 20: ...nto the sensor head until it is firmly seated in the correct position Tighten the securing screw slightly 8 Mount the sensor again In doing so use PTFE tape to seal the thread of the sensor head Check...

Страница 21: ...sensor no longer functions Only turn the shaft Never turn by the sensor head Make sure that the cable is not twisted Hold the sensor head firmly in position and unscrew the sensor shaft from the senso...

Страница 22: ...e one O ring on either side of the process connection Always replace the two O rings 5 a Release the O ring item a from the guide in the sensor shaft 6 Pull the O ring over the lamp and detector unit...

Страница 23: ...g product was ordered or delivered As an ISO certified company and also due to legal regulations Endress Hauser is obliged to follow certain procedures when handling any returned products that have be...

Страница 24: ...onfigurator on the product page www endress com cyh112 Technical Information TI00430C 11 Technical data 11 1 Input 11 1 1 Measured variable VIS NIR absorption 11 1 2 Measuring range 0 to 3 AU 0 to 6 O...

Страница 25: ...auser 25 11 3 Process 11 3 1 Process temperature 0 to 90 C 32 to 194 F continuous Max 130 C 266 F for 2 hours 11 3 2 Process pressure Max 10 bar 150 psi absolute at 20 C 68 F For version with Tri Clam...

Страница 26: ...ensor shaft Stainless steel 316L O ring EPDM Cable connector ends Nickel plated brass 11 4 4 Process connections Depends on version Thread G1 NPT 1 thread Tri Clamp 2 2 5 3 VARIVENT N DN40 125 11 4 5...

Страница 27: ...6 M Maintenance schedule 17 Measured variable 24 Measuring principle 8 Measuring range 24 Measuring system 11 Mounting angles 12 Mounting the sensor 14 N Nameplate 9 O Occupational safety 6 Operationa...

Страница 28: ...www addresses endress com 71330142 71330142...

Отзывы: