Technical data
Micropilot FMR67B HART
68
Hauser
T
a
T
p
P2
P1
P3
P4
P5
A0048826
19
Plastic housing material; process temperature range: 0 to +80 °C (+32 to +176 °F) with CSA C/US
approval
P1 = T
p
: 0 °C (+32 °F) | T
a
: +76 °C (+169 °F)
P2 = T
p
: +76 °C (+169 °F) | T
a
: +76 °C (+169 °F)
P3 = T
p
: +80 °C (+176 °F) | T
a
: +75 °C (+167 °F)
P4 = T
p
: +80 °C (+176 °F) | T
a
: 0 °C (+32 °F)
P5 = T
p
: 0 °C (+32 °F) | T
a
: 0 °C (+32 °F)
Process temperature range:
–40 to +150 °C (–40 to +302 °F)
T
a
T
p
P2
P1
P3
P4
P5
A0032024
20
Plastic housing material; process temperature range: –40 to +150 °C (–40 to +302 °F)
P1 = T
p
: –40 °C (–40 °F) | T
a
: +76 °C (+169 °F)
P2 = T
p
: +76 °C (+169 °F) | T
a
: +76 °C (+169 °F)
P3 = T
p
: +150 °C (+302 °F) | T
a
: +25 °C (+77 °F)
P4 = T
p
: +150 °C (+302 °F) | T
a
: –40 °C (–40 °F)
P5 = T
p
: –40 °C (–40 °F) | T
a
: –40 °C (–40 °F)
In the case of devices with plastic housings and CSA C/US approval, the selected
process temperature range of
–40 to +150 °C (–40 to +302 °F)
is limited to
0 to +150 °C (+32 to +302 °F).
Process temperature range:
0 to +150 °C (+32 to +302 °F), with CSA C/US approval
T
a
T
p
P2
P1
P3
P4
P5
A0048826
21
Plastic housing material; process temperature range: 0 to +150 °C (+32 to +302 °F) with CSA C/US
approval
P1 = T
p
: 0 °C (+32 °F) | T
a
: +76 °C (+169 °F)
P2 = T
p
: +76 °C (+169 °F) | T
a
: +76 °C (+169 °F)
P3 = T
p
: +150 °C (+302 °F) | T
a
: +25 °C (+77 °F)
P4 = T
p
: +150 °C (+302 °F) | T
a
: 0 °C (+32 °F)
P5 = T
p
: 0 °C (+32 °F) | T
a
: 0 °C (+32 °F)
Process temperature range:
–40 to +200 °C (–40 to +392 °F)