Micropilot FMR67B HART
Installation
Hauser
21
5.3.9
Changing the installation position of the display module
The installation position of the display can be changed in the case of the dual
compartment housing, L-form.
A0048401
L
WARNING
Supply voltage switched on!
Risk of electric shock and/or explosion!
‣
Switch off the supply voltage before opening the measuring device.
1 →
2 →
3 →
0.7 Nm
4 mm
2 x
1.
2.
2.
1.
A0046831
A0046832
A0046833
‣
If fitted: release the screw of
the cover lock for the display
cover using the Allen key.
‣
Unscrew the display cover and
check the cover seal.
‣
Press the release mechanism,
remove the display module.
4 →
5 →
6 →
0.7 Nm
4 mm
2 x
1.
2.
A0046834
A0046923
A0046924
‣
Release the connection.
‣
If fitted: release the screw of
the cover lock for the
connection compartment cover
using the Allen key.
‣
Unscrew the connection
compartment cover, check the
cover seal. Screw this cover
onto the electronics
compartment instead of the
display cover. If fitted: tighten
the screw of the cover lock
using the Allen key