Micropilot FMR67B HART
Diagnostics and troubleshooting
Hauser
47
Diagnostic
number
Short text
Remedy instructions
Status
signal
[from the
factory]
Diagnostic
behavior
[from the
factory]
942
In safety distance
1. Check level
2. Check safety distance
3. Reset self holding
S
Warning
952
Foam detected
Check process conditions
C
Warning
968
Level limited
1. Check level
2. Check limit parameters
S
Warning
1)
Diagnostic behavior can be changed.
11.9 Event logbook
11.9.1 Event history
A chronological overview of the event messages that have occurred is provided in the
Event logbook
submenu. This submenu only exists if operating via the local display. In the
case of operation via FieldCare, the event list can be displayed with the "Event List /
HistoROM" functionality of FieldCare.
Navigation:
Diagnostics → Event logbook → Event list
A maximum of 100 event messages can be displayed in chronological order.
The event history includes entries for:
• Diagnostic events
• Information events
In addition to the operating time when the event occurred, each event is also assigned a
symbol that indicates whether the event has occurred or is finished:
• Diagnostic event
• : Occurrence of the event
• : End of the event
• Information event
: Occurrence of the event
Calling up and closing the remedial measures
1. Press .
The message for the remedial measures for the selected diagnostic event opens.
2. Press + simultaneously.
The message about the remedial measures closes.
11.9.2 Filtering the event logbook
Filters can be used to determine which category of event messages is displayed in the
Event list
submenu.
Navigation: Diagnostics → Event logbook
Filter categories
• All
• Failure (F)
• Function check (C)