Micropilot FMR62
38
V. 4, Rev. 2, 19-03-2018
Hauser
Vibration resistance
DIN EN 60068-2-64 IEC 60068-2-64 at 5 to 2
000 Hz: 1.5 (m/s
2
)
2
/Hz
A limited vibration resistance applies according
DIN EN 60068-2-64/IEC 60068-2-64 at 5 to 2
000 Hz of 0.39 (m/s
2
)
2
/Hz,
if the following order features are selected in the product structure at the same time:
• 040 ("Housing"): B ("GT18 two-chamber, 316L)
and
• 090 ("Seal"):
– A6 ("FKM Viton GLT, –40 to 200 °C (–40 to 392 °F), incl. gas-tight feedthrough")
or
– C2 ("FKM Viton GLT, –40 to 200 °C (–40 to 392 °F), incl. gas-tight feedthrough")
or
– F6 ("FKM Viton GLT, –40 to 200 °C (–40 to 392 °F), incl. gas-tight feedthrough")
Electromagnetic
compatibility (EMC)
Electromagnetic compatibility in accordance with all of the relevant requirements outlined in the
EN 61326 series and NAMUR Recommendation EMC (NE 21). For details, please refer to the
Declaration of Conformity
.
A normal device cable suffices if only the analog signal should be used. Use a shielded cable for
digital communication (HART/ PA/ FF).
Maximum measured error during EMC testing: <0.5 % of the span. By way of derogation, for devices
with a plastic housing and see-through lid (integrated display SD02 or SD03) the measured error can
be up to 2 % of the span in the event of strong electromagnetic radiation in the 1 to 2 GHz frequency
range.
4)
Available for download at www.de.endress.com.