Memosens Wave CAS80E
Technical data
Hauser
37
13
Technical data
13.1 Input
Measured variable
• CODeq
• BODeq (mg/l)
• TOCeq (mg/l)
• TSS (mg/l)
• TU (FAU)
• APHA Hazen
(TU compensated/True Color or TU uncompensated/Apparent Color)
• SAK
• Nitrate NO3-N (mg/l)
Measuring range
The measuring range that can actually be achieved can depend on the composition of the
water matrix and the application. The data apply for homogeneous media.
Wastewater treatment plant inlet
Measured
variable
2 mm (0.08 in) gap
10 mm (0.4 in) gap
50 mm (1.97 in) gap
TSS
0 to 10
000 mg/l
0 to 2
000 mg/l
0 to 400 mg/l
SAC
0 to 1
000 1/m
0 to 200 mg/l
0 to 40 mg/l
CODeq
0 to 20
000 mg/l
0 to 4
000 mg/l
0 to 800 mg/l
BODeq
0 to 5
000 mg/l
0 to 1
000 mg/l
0 to 200 mg/l
Wastewater treatment plant outlet
Measured
variable
2 mm (0.08 in) gap
10 mm (0.4 in) gap
50 mm (1.97 in) gap
Turbidity
0 to 4
000 FAU
0 to 800 FAU
0 to 160 FAU
TSS
0 to 5
000 mg/l
0 to 1
000 mg/l
0 to 200 mg/l
SAC
0 to 1
000 1/m
0 to 200 1/m
0 to 40 1/m
CODeq
0 to 3
000 mg/l
0 to 600 mg/l
0 to 120 mg/l
TOCeq
0 to 1
200 mg/l
0 to 240 mg/l
0 to 48 mg/l
BODeq
0 to 450 mg/l
0 to 90 mg/l
0 to 18 mg/l
Nitrate NO3-N
0 to 2
500 mg/l
0 to 500 mg/l
0 to 100 mg/l
APHA Hazen true 0 to 12
500 Hazen
0 to 2
500 Hazen
0 to 500 Hazen
APHA Hazen
apparent
0 to 12
500 Hazen
0 to 2
500 Hazen
0 to 500 Hazen
1)
A minimum path length of 25 mm (0.98 in) is required in US Standard Methods 2120C (Single
Wavelength Method) 23. Edition
1)
eq = equivalent
2)
According to US Standard Methods 2120C (Single Wavelength Method) 23. Edition
3)
SAK
254
According to DIN ISO 38404-3