Memocheck CYP02D
Product description
Hauser
13
If the ambient temperatures shown above are not exceeded no invalid temperatures for the
particular temperature class will occur at the simulator.
7
Product description
Memocheck CYP02D consists of two interconnected sensor plug-in heads which are
independent of one another, and which each supply one specific parameter. You can simulate
the following values depending on your order:
Simulation values output for pH glass and pH ISFET
Fixed simulation value
Temperature
pH glass
4
25 ˚C (77 ˚F)
pH ISFET
7
60 ˚C (140 ˚F)
Simulation values output for pH glass and pH glass
Fixed simulation value
Temperature
pH glass
4
25 ˚C (77 ˚F)
pH glass
7
60 ˚C (140 ˚F)
Simulation values output for pH glass and ORP
Fixed simulation value
Temperature
pH glass
4
25 ˚C (77 ˚F)
ORP potential
750 mV
90 ˚C (194 ˚F)
Simulation values output for pH glass and oxygen
Fixed simulation value
Temperature
pH glass
4
25 ˚C (77 ˚F)
Oxygen
60 nA
60 ˚C (140 ˚F)
1)
Sensor current
Simulation values output for pH glass and conductive measurement of conductivity
Fixed simulation value
Temperature
pH glass
4
25 ˚C (77 ˚F)
Conductivity
20 µS/cm
60 ˚C (140 ˚F)